30
out. The icon of each feature you
lock out will appear. You may lock
out one or all features.
4. Press the
SET button
when finished to
set lock outs.
Unlocking the Lock-Outs:
Repeat Steps 1 and 2 on how to set the
Lock Out feature. Select the operating
features you wish to unlock and press
the corresponding unlock and press the
corresponding button. The feature’s icon
will disappear from the LCD. Press the SET
button when finished.
Timer
The timer is used to monitor the number of
treatment hours a patient receives.
1. Turn power off from the POWER
ON/OFF switch.
2. Press and hold the PAUSE button and
the POWER ON switch simultaneously.
3. The display represents the number
of hours the device has been used
(to the closest hour).
4. To reset the timer to 0 hours for the
next patient, press and hold the
– button until 0 hours is displayed. The
device will now begin counting usage
hours from 0.
The timer counts to a maximum of 999
hours after which is must be reset.
6. Conversion
Replacing/Changing Finger
Actuators
The
ARTROMOT®-F
can be converted
easily:
1. There are 3 medium finger actuators
(black) and one small finger actuator
(blue). Additional finger actuators in
small, medium or large size may be
ordered separately. See the Ordering
Information chart for details.
2. To replace finger actuators, detach the
black bar (6) containing the finger actu-
ators by removing both fastening screws
with a Phillips screwdriver (20).
3. With the bar detached, slide the finger
actuators off the bar and replace or rear-
range them for left or right hand set up.
4. When changing from right to left hand,
or vice versa, remove the finger actu-
ators and rearrange them so that the
short finger actuator (blue) is always in
position to attachment to the smallest
finger.
7. Maintenance
– Always unplug the device before clea-
ning.
– The
ARTROMOT®-F
can be
wiped
clean
with disinfectant and thus fulfills the
hygienic requirements for medical equip-
ment.
– Only wipe the motion device itself with a
moist cloth.
m
PRECAUTION!
No fluids may be allowed to
enter the casing or the hand-held pro-
gramming unit.
– Protect the motion device from intensi-
ve UV light (sunlight)
ATTENTION WARNING
Complying with the following in structions
will help reduce unnec essary wear and
damage to the
ARTROMOT®-F
.
1.
ALWAYS
use the hand-held program-
ming unit to run the device
to adjust the position of the Drive Bar
and Finger Actuators.
2.
DO NOT
force the actuators to move
by hand by pushing on the Drive Bar.
This action breaks the gear in trans-
mission and requires the gear motor
to be replaced.
This action also bre-
aks the Arm Drive Gear Support Pins
and requires that each damaged
Bridge Arm assembly be replaced.
(fig.23/24)
MOT-301-DE_USA-GB_Rev_03-04-01-2017_DV3.indd 30
17.01.17 10:10
Содержание ARTROMOT-F
Страница 2: ...MOT 301 DE_USA GB_Rev_03 04 01 2017_DV3 indd 2 17 01 17 10 10 ...
Страница 4: ...4 Figures 1 2 3 4 5 6 7 8 Abbildungen MOT 301 DE_USA GB_Rev_03 04 01 2017_DV3 indd 4 17 01 17 10 10 ...
Страница 5: ...5 9 10 11 12 13 14 16 15 MOT 301 DE_USA GB_Rev_03 04 01 2017_DV3 indd 5 17 01 17 10 10 ...
Страница 21: ...21 MOT 301 DE_USA GB_Rev_03 04 01 2017_DV3 indd 21 17 01 17 10 10 ...
Страница 37: ...37 MOT 301 DE_USA GB_Rev_03 04 01 2017_DV3 indd 37 17 01 17 10 10 ...
Страница 38: ...38 Notes MOT 301 DE_USA GB_Rev_03 04 01 2017_DV3 indd 38 17 01 17 10 10 ...
Страница 39: ...39 Notes MOT 301 DE_USA GB_Rev_03 04 01 2017_DV3 indd 39 17 01 17 10 10 ...