background image

www.DJOglobal.eu

3. 

Connect System

Connect cuffs to the tubing that is attached to the pump.

Intended Use/Indications

The VenaFlow Elite System is intended as a prophylaxis for deep vein thrombosis (DVT).

Contraindications

The VenaFlow Elite System should not be used by persons with known or suspected deep vein thrombosis, severe congestive heart failure, pulmonary edema, thrombophlebitis, severe arteriosclerosis, or active infection. Do 
not use on extremities which are not sensitive to pain, where cuff will interfere with gangrene, on patients with vein ligation or recent skin grafts, or extreme deformity of the leg. Do not use the VenaFlow Elite System where 
increased venous or lymphatic return is undesirable.

Warning and Precautions

•  The VenaFlow Elite cuffs are designed for single patient use only.
•  Risk of explosion if used in the presence of flammable anaesthetics.
•  The equipment or system should not be used adjacent to or stacked with other equipment and, if adjacent or stacked use is necessary, the equipment or system should be observed to verify normal operation in the 

configuration in which it will be used.

•  Medical Electrical Equipment needs special precautions regarding EMC. Portable and Mobile RF Communications Equipment can be affected by other Medical Electrical Devices.
•  Cuffs used in combination with warming devices may cause skin irritation. Regularly check for patient comfort, compliance and skin irritation.
•  Special attention should be given to those patients with neuropathies or tissue viability problems (i.e. diabetes, arterial or venous insufficiencies).
•  To prevent extremity compartment syndrome, special attention should be given to patients who are positioned in the supine lithotomy position for extended lengths of time. This includes patients with or without 

compression cuffs.

•  If you experience pain, swelling, sensation changes or any unusual reactions while using this device, consult your medical professional immediately.

4. 

Standard Rapid Mode Operation

  Press ON/OFF/RESET button. 

  Cuff inflation alternates between the two legs every 30 

seconds.

  Distal aircell inflates first and the proximal aircell follows. 

  Pressures settle at 45mmHg ± 15%. (Foot Cuff peak ~130 

mmHg.)

  After six seconds the cuff deflates. 

  Alarm will sound if the system is not functioning properly.

5. 

S Mode Operation

  When powered on, the VenaFlow Elite System will default 

to standard rapid inflation mode. If a slower inflation is 
desired, select the S mode operation 

 button. 

  Once selected, both cuffs will inflate simultaneously once 

per minute.

  Both cuffs will inflate and compress for a total of 10 seconds. 

  While in S mode the icon 

 will appear in the top center 

of the graphical display. The device will remain in S mode 
until user presses the S mode operation button again. 
This button allows user to toggle between standard rapid 
inflation mode and S mode.

6. 

Locking in Standard Rapid   

or S Mode

  To lock in standard rapid mode, make sure the device is in 

standard rapid mode, then press and hold the S mode 

 

button for 5 seconds. Do the same to unlock it.

  To lock in S mode, make sure the device is in S mode and 

press and hold the S mode button 

 for 5 seconds. Do 

the same to unlock it. 

  Once the system is locked in either standard or S mode, the 

lock symbol 

 will appear on the screen.

Check Tubes Alarm

(Kinked or Disconnected Tube)

Call For Service Alarm

(Ship back to DJO for repair)

Alarm Reset

  To reset any alarm, press the center 

ON/OFF/RESET   button and 
take steps to correct the alarm if 
necessary. 

  Note: Pressing ON/OFF/RESET 

following alarm only 

resets

 the alarm.  

The device will continue to run and will 
not shut down. Alarm will continue at 
next cycle if not corrected.

Отзывы: