Assembly 

Housing

Maintenance

n

Clearing a clogged ferrule assembly

If fiber happens to break off inside the ferrule assembly,
Clean-Out Wire and a Clean-Out Wire Guide is supplied to
assist in removing any debris.  Disconnect the patch cable
from the Divot® Module by rotating the LiteLOCK® dial
slightly to the left and pull the connector out from the
interface.  Remove the end cap, OCC cartridge and sleeve.
Pull the ferrule assembly out of the bulkhead housing.
Push the ferrule assembly into the 2.5mm interface of the
Clean-out Wire Guide.  Insert a piece of clean-out wire into
the small hole on the end of the Clean-out Wire Guide and
slowly push forward to clean out any broken fiber or debris.
Reassemble the Divot® Module and insert the connector
with the red boot or red band from the patch cable into the
Universal Bulkhead Housing and secure by  rotating the
LiteLOCK® dial to the right.  Do not overtighten.

n

Cartridge Replacement

The OCC cartridge is filled with optical coupling compound and under normal use will typically
produce a minimum of 500 insertions.  The cartridge needs to be replaced when you start to
experience inconsistent launches. In some cases, especially over prolong use, debris can build
up in the cavity on the tip of the Ferrule Assembly which will cause inconsistency in the
connection.  Before replacing the cartridge, clean out the ferrule assembly.  Please refer to page
2 for instructions on how to replace (load) the Optical Compound Cartridge.

Insert 

Clean-out Wire

n

Connecting and Disconnecting the patch cable for cleaning

Occasionally you may have to clean the connector on the patch cable (identified with a red boot
or band below the boot) and/or both ends of the ferrule assembly inside the Divot® Module.  The
ferrule assembly can be removed from the bulkhead housing and cleaned or immersed in
alcohol.  To remove the Divot® Module from the Patch Cable simply loosen the LiteLOCK® dial
by rotating it slightly to the left and pull the connector out from the interface.  To reconnect the
patch cable, insert the connector marked with a red boot or red band below the connector boot
into the Universal Bulkhead Housing.  Push the ferrule of the connector into the 2.5mm interface
and secure by rotating the LiteLOCK® dial to the right.  Do not overtighten.  The LiteLOCK® dial
can also be unscrewed from the bulkhead housing in order to clean or replace the zirconia
sleeve.  Over time, especially if inserting non-cleaved fiber, the patch cable mated to the Divot®
module will wear and may need to be replaced.  Additional cables are available as accessories.

Ferrule

Assembly

Phos. Bronze

Sleeve

OCC

Cartridge

End Cap

Universal 

LiteLOCK® Dial

Zirconia 

Sleeve

Specifications & Accessories

n

Specifications

Fiber type:  Singlemode 9/125µm, Multimode 62.5/125µm or Multimode 50/125µm
Cable length:  1 meter
Number of insertions (Typical):  3000 min. (500 min. per cartridge, 6 cartridges included)
Insertion loss (Typical):  < 0.5 dB (uncleaved fiber < 0.8dB)
Back Reflection (Typical): < 45 dB
Operating temp.:  -10° C to +40° C

n

Accessories

DVT-RC3 

Divot® Replacement OCC Cartridges (3 pack)

DVT-RC12 

Divot® Replacement OCC Cartridges (12 pack)

COKS-40

40 pc. Vial of Clean-Out Wire

DVT-RP4000

Divot® Replacement Sleeve Set (1 Zirconia, 1 Phos. Bronz)

DVT-RP4100

Divot® Replacement Ferrule Assembly

---------

Divot® Replacement Patch Cable Assembly

---------

Dead Zone Eliminator® Launch Box (Fiber Spool) for use with OTDR’s

3

Содержание DVTK

Страница 1: ...Tester n Divot Bare Fiber Testing Device DVTK The Divot Bare Fiber Adapter Tester is a device that quickly connects unterminated bare fiber to OTDR s or other fiber optic test equipment Simply strip the fiber cleave and insert into the Divot Module In an instant the fiber is connected to your OTDR with low loss and low reflectance Items not included which may be required n Optical Strippers n Scri...

Страница 2: ...ble assembly If a connection is not made on the first insertion pull the fiber back slightly into the module and push forward again If using uncleaved fiber and are unable to get a successful connection after a few attempts remove the fiber and rebreak or cut the fiber again n Proceed with testing OTDR Divot Module Fiber Getting Started Instructions Screw on the End Cap until it stops The Divot Mo...

Страница 3: ...lean the connector on the patch cable identified with a red boot or band below the boot and or both ends of the ferrule assembly inside the Divot Module The ferrule assembly can be removed from the bulkhead housing and cleaned or immersed in alcohol To remove the Divot Module from the Patch Cable simply loosen the LiteLOCK dial by rotating it slightly to the left and pull the connector out from th...

Страница 4: ... manufactured by Fiber Plus International USA Products protected by one or more of the following US and international patents and patents pending Patent nos 5574813 5751874 5761359 5769698 D409154 5781285 5818993 6503422 6095694 7014370 Fiber Plus intl reserves the right to improve enhance or modify the features and specifications of products without prior notification 2006 2014 Fiber Plus Interna...

Отзывы: