background image

Before you begin:

Load Cartridge

An OCC cartridge needs to be loaded into the Divot® Module before
use.  Six sealed cartridges are included in the case.

n

Loading an OCC Cartridge

You do not need to remove the Divot® module from the patch cable
when loading a  cartridge.

End Cap

1

Unscrew and remove the 

End Cap

to expose the split sleeve.

2

Remove the two plastic tabs (located on each end) from one of the
OCC cartridges included.  Leave remaining cartridge sealed until
ready for use.

l

Tabs

l

l

3

After removing the tabs from the OCC cartridge, push the larger end
into the sleeve.

4

n

Connect patch cable to OTDR or test equipment.

n

Break out a few inches of the fiber to be tested down to 250µm buffer.

Strip fiber down to 125µm cladding.  Cleave fiber leaving a minimum of 3/4”
of fiber.  Clean or wipe off fiber if needed.

n

Insert your prepared bare fiber end into the round opening on the end of

the OCC cartridge protruding from the Divot® module.

The bare fiber will be stopped by the attached patch cable assembly.  If a connection is not
made on the first insertion, pull the fiber back slightly into the module and push forward again.
If using uncleaved fiber and are unable to get a successful connection after a few attempts,
remove the fiber and rebreak or cut the fiber again.

n

Proceed with testing.

OTDR

Divot® Module

Fiber

Getting Started

Instructions

Screw on the End Cap until it stops.  The Divot® Module is now
loaded and ready for use.  Place the dust cap included in the case
(not shown) onto the end of the cartridge protruding from the Divot®
module when storing the unit.

5

Insert Divot® module into the Desktop Mount/Holder by removing the
top half of the holder.  Place the Divot® module into the cut out
section on the bottom half of the holder.  Place the top section over
the Divot® module.  The top section is held in place by magnets.
There should be a slight gap between the top and bottom section of
the Desktop Mount/Holder.

6

Caution: Always wear protective eyewear when working with bare or active fiber.  Never look
into the end of the Divot® Module when patch cable is connected to active test equipment.

Questions? Please contact Customer Service at 1-866-640-3468 or visit www.fiberplus.com.

?

2

Содержание DVTK

Страница 1: ...Tester n Divot Bare Fiber Testing Device DVTK The Divot Bare Fiber Adapter Tester is a device that quickly connects unterminated bare fiber to OTDR s or other fiber optic test equipment Simply strip the fiber cleave and insert into the Divot Module In an instant the fiber is connected to your OTDR with low loss and low reflectance Items not included which may be required n Optical Strippers n Scri...

Страница 2: ...ble assembly If a connection is not made on the first insertion pull the fiber back slightly into the module and push forward again If using uncleaved fiber and are unable to get a successful connection after a few attempts remove the fiber and rebreak or cut the fiber again n Proceed with testing OTDR Divot Module Fiber Getting Started Instructions Screw on the End Cap until it stops The Divot Mo...

Страница 3: ...lean the connector on the patch cable identified with a red boot or band below the boot and or both ends of the ferrule assembly inside the Divot Module The ferrule assembly can be removed from the bulkhead housing and cleaned or immersed in alcohol To remove the Divot Module from the Patch Cable simply loosen the LiteLOCK dial by rotating it slightly to the left and pull the connector out from th...

Страница 4: ... manufactured by Fiber Plus International USA Products protected by one or more of the following US and international patents and patents pending Patent nos 5574813 5751874 5761359 5769698 D409154 5781285 5818993 6503422 6095694 7014370 Fiber Plus intl reserves the right to improve enhance or modify the features and specifications of products without prior notification 2006 2014 Fiber Plus Interna...