WARNING:  If you plan to perform an Carbon 

Monoxide sensor calibration by using ambient air make 
sure you are far from sources of pollution like cars, 
trucks, boats and that you let cootwo ambient for at 
least 5 minutes before performing the calibration as the 
gas flow within the device is reduced and a shorter 
period might produce incorrect results. 

We recommend you perform the ZERO point calibration 

frequently. 

8.2.2 

SPAN  Calibration 

The procedure to perform the SPAN calibration is similar 

to the one used for the ZERO point calibration; the most 
notable difference is that you will need to use a certified 
calibration gas containing 20ppm of Carbon Monoxide. 

8.3 

CO Sensor Replacement 

You can replace  the Carbon Monoxide sensor yourself 

or send your cootwo back to us for service and sensor 
replacement. 

If you want to replace the CO sensor yourself you will 

need first to open cootwo by removing the 3 screws on the 
back of the unit and the large O-ring on the tank adapter. 

Then unplug the CO sensor from the board and replace 

it with the new one and close the unit. 

WARNING: the Carbon Monoxide sensor contains 

chemicals which can cause burns if they come in 
contact with eyes, skin and other body parts. Dispose 
of the Carbon Monoxide sensor according to local 
regulations. 

After you have physically replaced the sensor you will 

need to properly program the analyzer; to do so go to the 
CO SEN sub-menu, scroll down to the CHANGE sub-menu 
and enter it. 

 

Now, PUSH the button to confirm 

that you have replaced the Carbon 
Monoxide  sensor and cootwo will 
acknowledge it and turn off after a few 
seconds. 

 

Please note that when you replace  the  Carbon 

Monoxide sensor you will need to calibrate the new sensor; 
first you will need to perform the ZERO point calibration 
using a gas containing 0 ppm CO and then you will need to 
perform the SPAN calibration. 

WARNING: After replacing a Carbon Monoxide 

sensor you will need to make sure that cootwo is 
properly powered (battery is almost full) and wait at 
least 4 hours before you can proceed with the CO 
sensor calibration. Performing  a CO sensor calibration 
too soon after a sensor replacement might produce 
incorrect results.

  

9  BATTERY 

cootwo  is equipped with a  user replaceable  LIR 10440 

AAA 3.7V rechargeable battery that can be easily recharged 
using the provided USB cable. We strongly recommend you 
keep the battery properly charged all times. 

WARNING: If you do not keep the battery properly 

charged the battery might die off and you might not be 
able to recharge it all. 

WARNING: If you let the battery become very empty 

or even die, it might take several hours to recharge it. 

WARNING: If you let the battery die, the Carbon 

Monoxide sensor might loose power. When the Carbon 
Monoxide looses power it will loose its bias. Once you 
re-apply power to the Carbon Monoxide sensor it might 
need several hours to recover its original status. 

10 WARRANTY 

DiveNav, Inc. (the “Company”) warranty obligations for 

cootwo (the “Product”) are limited to the terms set forth in 
this document as well as  those set forth in the Warranty 
section of the company website.  

Company warrants the Product against substantial 

defects in material and workmanship under normal use for a 
period of  one (1) year from the date of original purchase 
(the “Limited Warranty”).  

This Limited Warranty is non-transferable and covers 

only the original purchaser.  

An original or copy of the sales receipt from the original 

sales retailer or Company website is required to validate the 
warranty coverage.  

This Limited Warranty does not cover the battery and 

products purchased through non-authorized  dealers. 
Warranty claims must be made directly to the Company. 

10.1 

Limitation of Liability 

IN NO EVENT SHALL COMPANY, NOR ITS 

AFFILIATES, NOR ITS EMPLOYEES, OFFICERS, 
DIRECTORS, AGENTS, CONTRACTORS, 
DISTRIBUTORS, SUPPLIERS OR ASSIGNS (THE 
“RELEASED PARTIES”)   BE LIABLE TO YOU OR ANY 
THIRD PARTY FOR ANY INDIRECT, INCIDENTAL, 
CONSEQUENTIAL, SPECIAL, EXEMPLARY OR PUNITIVE 
DAMAGE ARISING OUT OF THE USE OF THE 
PRODUCT, INCLUDING, WITHOUT LIMITATION, 
PROPERTY DAMAGE, LOSS OF VALUE OF THE 
PRODUCT, DEATH, INJURY OR ANY OTHER DAMAGES 
THAT MAY BE CAUSED TO YOU, YOUR FAMILY, HEIRS, 
ESTATE OR ASSIGNS. NOTWITHSTANDING ANY 
DAMAGES THAT YOU MIGHT INCUR FOR ANY REASON 
WHATSOEVER (INCLUDING, WITHOUT LIMITATION, ALL 
DAMAGES REFERENCED HEREIN) AND ALL DIRECT 
OR GENERAL DAMAGES IN  CONTRACT, TORT 
(INCLUDING NEGLIGENCE AND OTHERWISE), THE 
ENTIRE  LIABILITY OF THE COMPANY SHALL BE 
LIMITED TO THE AMOUNT ACTUALLY PAID BY YOU 
FOR THE PRODUCT. SOME STATES DO NOT ALLOW 
THE EXCLUSION OR LIMITATION OF INCIDENTAL OR 
CONSEQUENTIAL DAMAGES, SO THE ABOVE 
LIMITATIONS OR EXCLUSIONS MAY NOT APPLY TO 
YOU. 

10.2 

Exclusions and Limitations 

Company’s Limited Warranty applies only to the 

Product. The warranty does not apply to any non-Company 
products, even if packaged or sold with the Product. This 
warranty does not apply to damage caused by abuse, 
misuse, accident, tampering, force majeure, or modifications 
of the Product. 

Содержание Cootwo

Страница 1: ...Dual gas Oxygen and Carbon Monoxide SMART Analyzer User Manual ...

Страница 2: ...SENSOR CALIBRATION 5 7 2 O2 SENSOR REPLACEMENT 6 8 CARBON MONOXIDE SENSOR 6 8 1 CO SENSOR BUMP TEST 6 8 2 CO SENSOR CALIBRATION 6 8 2 1 ZERO POINT CALIBRATION 6 8 2 2 SPAN CALIBRATION 7 8 3 CO SENSOR REPLACEMENT 7 9 BATTERY 7 10 WARRANTY 7 10 1 LIMITATION OF LIABILITY 7 10 2 EXCLUSIONS AND LIMITATIONS 7 10 3 REMEDY 8 10 4 WARRANTY DISCLAIMER 8 10 5 OBTAINING WARRANTY SERVICE 8 11 FCC 8 11 1 FCC CO...

Страница 3: ...manual WARNING cootwo includes an Oxygen sensor that must be calibrated before each use An improper calibration may result in the use of an incorrect breathing gas mix which could lead to serious personal injury and even death WARNING Do not use cootwo if you are not Nitrox certified WARNING Do not use cootwo if it is not calibrated WARNING Keep cootwo away from heat sources WARNING Do not expose ...

Страница 4: ... to limit the gas flow to few liters per minute If you have ordered the Low Pressure Inflator adapter then you can connet your cootwo directly to the low pressure hose Once you are ready PUSH the activation button to start the Oxygen sensor calibration As mentioned above the stand alone manual Oxygen calibration is limited to air it is time based and it takes 1 minute Now that cootwo is calibrated...

Страница 5: ...r in stand alone mode or via the My Nitroxbuddy app Please note that it is YOUR responsibility to set the sensor life correctly The analyzer only provides you a reminder but it has absolutely no knowledge if the lifetime you entered is correct for the O2 sensor you are using 7 1 O2 Sensor Calibration As indicated earlier with cootwo you can perform an Oxygen sensor calibration either with the stan...

Страница 6: ...r costs and procedures 8 1 CO Sensor Bump Test It is good practice to periodically verify if the Carbon Monoxide sensor is working properly The simplest way to do so is to perform a bump test using your breath According to a paper from Ryter and Choi the exhaled Carbon Monoxide value of non smokers could be about 3 ppm while the mean value of exhaled Carbon Monoxide for smokers was 17 ppm Before y...

Страница 7: ...end you keep the battery properly charged all times WARNING If you do not keep the battery properly charged the battery might die off and you might not be able to recharge it all WARNING If you let the battery become very empty or even die it might take several hours to recharge it WARNING If you let the battery die the Carbon Monoxide sensor might loose power When the Carbon Monoxide looses power...

Страница 8: ...problem 4 Provide a telephone number and e mail address at which you can be reached during normal business hours All of the above information is required before a warranty claim will be accepted Once Company validates your claim we will issue you an RMA Return Material Authorization number and shipping information 11 FCC 11 1 FCC Compliance Statement FCC ID RSY2015COOTWO This device complies with ...

Отзывы: