DiveNav Cootwo Скачать руководство пользователя страница 5

 

cootwo User Manual – Rev. 20170512                               ©2017 DiveNav, Inc.  – www.divenav.com                                          

         P a g e     5 

If you try to access the Main Settings menu while the 

analyzer is connected to a smartphone, the analyzer will 
display  a warning  screen  and  will give you the option to 

disconnect from the smartphone. 

 
This  image  on the right shows the 

Main Settings menu. 

 
This menu contains 5 sub-menus: 

BACK, O2 CAL, SENSOR, LIGHT and 
DATE.  

To access a sub-menu first position the > pointer on the 

line you want to select by as many PUSH as required then 
enter the sub-menu by a HOLD. 

If you HOLD when BACK is selected you will go back to 

the previous menu. In the case of the image above you will 
go back to the Normal Operating Mode. 

The S symbol on the top row indicates that you are in 

the settings operating mode. 

The O2 CAL sub-menu allows you to perform an 

Oxygen sensor calibration. 

The SENSOR sub-menu allows you to access settings 

for either the Oxygen sensor or the Carbon Monoxide 
sensor. 

The LIGHT sub-menu allows you to program intensity 

and duration of the backlight. 

The DATE sensor allows you to program date and time. 
Please note that the settings available in stand-alone 

mode are a sub-set of the full set of settings available with 
the My Nitroxbuddy app. 

See below the top level view of cootwo menus. 
 

7  OXYGEN SENSOR 

cootwo comes equipped with an electrochemical 

Oxygen sensor. The life of an Oxygen sensor is difficult to 
estimate as it might be affected by several factors such as 
usage, exposure to a gas with high content of Oxygen, 
temperature and humidity.  

Also, an Oxygen sensor does not die immediately but 

slowly degrades over time. 

We recommend to replace the factory installed Oxygen 

sensor every 12 months. But, if you are in the middle of a 
dive trip you can simply extend the life of the O2 sensor by 
simulating a replacement either in stand-alone mode or via 
the My Nitroxbuddy app.  

Please note that it is 

YOUR

  responsibility to set the 

sensor life correctly. The analyzer only provides you a 
reminder but it has absolutely no knowledge if the lifetime 
you entered is correct for the O2 sensor you are using. 

7.1 

O2 Sensor Calibration 

As indicated earlier, with cootwo you can perform an 

Oxygen sensor calibration either with the stand-alone unit or 
with the assistance of a smartphone.  

To calibrate the Oxygen sensor while in stand-alone 

mode, first turn the unit ON. 

Then, while the unit is in the Normal Operating Mode, 

HOLD the button and enter the Main Settings menu.  

 
 
Now, PUSH the button to position 

the pointer on the O2 CAL sub-menu as 
per image on the right. 

 

 

Now, HOLD the button to enter the 

Oxygen Sensor calibration  sub-menu 
show in the image on the left. 

As  indicated earlier, HOLD enters the sub-menu 

selected by the pointer while a PUSH moves to the next line 
down. In this specific case, HOLD will go back to the 
previous menu while PUSH will start the calibration of the 
Oxygen sensor. 

 During an Oxygen sensor calibration it is very important 

to expose cootwo to a known source of gas flowing at a 
constant rate.  

WARNING: If you plan to perform an Oxygen sensor 

calibration by exposing cootwo to ambient air MAKE 
SURE cootwo acclimates for at least 5 minutes as the 
gas flow within the device is reduced and a shorter 
period might produce incorrect results. 

Now, position your cootwo in front of the valve of a 

scuba tank containing air and open the tank valve slowly so 
to limit the gas flow to few liters per minute. If you have 
ordered the Low Pressure Inflator adapter then you can 
connect your cootwo directly to the low pressure hose.  

Once you are ready PUSH the activation button to start 

the Oxygen sensor calibration and cootwo will alternate the 
2 screens below until the O2 calibration is completed. 

 
 
 
 
 
 
 
 
 

 

As mentioned above, the stand-alone,  manual  Oxygen  

calibration is limited to air, it is time based and it takes about 
one minute. 

Содержание Cootwo

Страница 1: ...Dual gas Oxygen and Carbon Monoxide SMART Analyzer User Manual ...

Страница 2: ...SENSOR CALIBRATION 5 7 2 O2 SENSOR REPLACEMENT 6 8 CARBON MONOXIDE SENSOR 6 8 1 CO SENSOR BUMP TEST 6 8 2 CO SENSOR CALIBRATION 6 8 2 1 ZERO POINT CALIBRATION 6 8 2 2 SPAN CALIBRATION 7 8 3 CO SENSOR REPLACEMENT 7 9 BATTERY 7 10 WARRANTY 7 10 1 LIMITATION OF LIABILITY 7 10 2 EXCLUSIONS AND LIMITATIONS 7 10 3 REMEDY 8 10 4 WARRANTY DISCLAIMER 8 10 5 OBTAINING WARRANTY SERVICE 8 11 FCC 8 11 1 FCC CO...

Страница 3: ...manual WARNING cootwo includes an Oxygen sensor that must be calibrated before each use An improper calibration may result in the use of an incorrect breathing gas mix which could lead to serious personal injury and even death WARNING Do not use cootwo if you are not Nitrox certified WARNING Do not use cootwo if it is not calibrated WARNING Keep cootwo away from heat sources WARNING Do not expose ...

Страница 4: ... to limit the gas flow to few liters per minute If you have ordered the Low Pressure Inflator adapter then you can connet your cootwo directly to the low pressure hose Once you are ready PUSH the activation button to start the Oxygen sensor calibration As mentioned above the stand alone manual Oxygen calibration is limited to air it is time based and it takes 1 minute Now that cootwo is calibrated...

Страница 5: ...r in stand alone mode or via the My Nitroxbuddy app Please note that it is YOUR responsibility to set the sensor life correctly The analyzer only provides you a reminder but it has absolutely no knowledge if the lifetime you entered is correct for the O2 sensor you are using 7 1 O2 Sensor Calibration As indicated earlier with cootwo you can perform an Oxygen sensor calibration either with the stan...

Страница 6: ...r costs and procedures 8 1 CO Sensor Bump Test It is good practice to periodically verify if the Carbon Monoxide sensor is working properly The simplest way to do so is to perform a bump test using your breath According to a paper from Ryter and Choi the exhaled Carbon Monoxide value of non smokers could be about 3 ppm while the mean value of exhaled Carbon Monoxide for smokers was 17 ppm Before y...

Страница 7: ...end you keep the battery properly charged all times WARNING If you do not keep the battery properly charged the battery might die off and you might not be able to recharge it all WARNING If you let the battery become very empty or even die it might take several hours to recharge it WARNING If you let the battery die the Carbon Monoxide sensor might loose power When the Carbon Monoxide looses power...

Страница 8: ...problem 4 Provide a telephone number and e mail address at which you can be reached during normal business hours All of the above information is required before a warranty claim will be accepted Once Company validates your claim we will issue you an RMA Return Material Authorization number and shipping information 11 FCC 11 1 FCC Compliance Statement FCC ID RSY2015COOTWO This device complies with ...

Отзывы: