Servicing

 DANGER SHOCK HAZARD

Possible shock hazard may occur if the main power supply cord for the control box is not unplugged from 

the wall outlet before any maintenance is performed on electrical components.

Actuators and Control Box

Cord and Socket Identification

Replacing Control Box

•  Unplug mains cable from the wall outlet

•   Unplug connections for all actuators 

 

and accessories

•  Separate the control box from the actuator

•  Reinstall the control box on the actuator

•   Plug all connections back into control box 

following color coding

•  Plug mains cable into wall outlet

•  Test function

 

WARNING

: When routing cables in the medical 

bed, precautions shall be taken to avoid squeezing 

between parts of the medical bed.

Replacing the AC Power Cord (mains cable)

•  Unplug mains cable from wall outlet

•  Unplug mains cable from the control box

•   Remove tie-wraps holding power cord in place 

on frame

•  Plug the new power cord into the control box

•   Route the new power cord in the same manner 

as the original cord, ensuring use of strain relief 

device at head end of bed

•  Replace tie-wraps to hold power cord to frame

•  Plug mains cable into wall outlet

•  Test function

 

WARNING

: When routing cables in the medical 

bed, precautions shall be taken to avoid squeezing 

between parts of the medical bed.

Replacing Actuators

•  Unplug mains cable from the wall outlet

•  Identify the actuator to replace

•   With the help of a second person, place bed on 

its side to remove the actuators

•  Unplug actuator cord from control box

•  Actuator is held in place by (2) clevis pins

•  Remove bowtie clips from clevis pins

•  Slide the clevis pins out of the holes

•  Replace the actuator

•   To reassemble bed, reverse previous steps, and 

make sure to:

  -   Assemble clevis pins as originally installed with 

bowtie clips

  -   Zip ties should be replaced, with cords in their 

original position and routing direction to the 

control box

•  Plug mains cable into wall outlet

•  Test function

 

DANGER

: Failure to replace bowtie clips on 

clevis pins at each end of actuator can result in 

harm to the user, operator and/or equipment.

 

WARNING

: When routing cables in the medical 

bed, precautions shall be taken to avoid squeezing 

between parts of the medical bed.

Bed End Panel 

(Green)

Hand Pendant 

(Red)

Head Section Motor 

(Black)

Foot Section Motor 

(Yellow)

Foot End Hi-Lo Motor 

(Blue)

Head End Hi-Lo Motor 

(White)

Battery Backup 

(Black)

Содержание Panacea 6300

Страница 1: ...d from Direct Supply Equipment Furnishings Please read this entire guide carefully and keep it for future reference This guide will provide you with instructions warnings warranty information and othe...

Страница 2: ...eadboard and Footboard Installation and Adjustment 16 Assist Device Assembly 17 3 Position Assist Device 18 Width Extension 19 Wall Guard 20 Bed Functions 21 Bed Mobilization and Stabilization 22 Batt...

Страница 3: ...s and Symbols NOTE Indicates a helpful tip CAUTION Indicates correct operating or maintenance procedures in order to prevent damage to or destruction of the equipment or other property WARNING Calls a...

Страница 4: ...ensure tubing does not become entangled kinked or severed during normal operation of the bed Warning This device includes assist devices Some states may prohibit assist devices or consider assist dev...

Страница 5: ...re cord does not become entangled or damaged during normal operation of the bed Keep all moving parts free from obstruction e g blankets sheets heating blankets pads tubing wiring etc which may get ta...

Страница 6: ...etainers inward Perform this procedure on both sides of the bed The bed is now at 37 wide and can be used to evacuate resident out of the facility WARNING The device should ONLY be used to transport r...

Страница 7: ...to clean if necessary Inspect the resident s bed controls looking for signs of damage where liquids could leak in Test the bed and its controls including resident lockout features to ensure the bed c...

Страница 8: ...ISO 7000 1641 Operating instructions Change Auto Contour position Recycle the item in accordance with local regulations Change hi lo position of mattress support platform Intertek s ETL listed mark f...

Страница 9: ...nto the control box of bed i e pendants and actuators while mains cable power cord is plugged into the wall outlet Any cords or tubing used on or with this bed MUST be routed and secured properly to e...

Страница 10: ...s and options not all accessories and options can be used at the same time Assumes Max Patient Resident Weight 582 lbs 264kg Assist Rail pair 35 lbs 16kg Assist Rail 4 70 lbs 32kg Headboard Footboard...

Страница 11: ...ame and assist rails are secured to frame to decrease risk of entrapment On March 10 2006 the U S Food and Drug Administration FDA released guidelines for reducing the risk of hospital bed entrapment...

Страница 12: ...System 5 Locking Casters Fall Mat Left and Right Side Pendant Connections Trapeze Free Standing or Bed Mounted Slat Deck Staff Control Panel Heavy Duty Casters Mattress Retainers Beige Powdercoated F...

Страница 13: ...endelenburg Reverse Trendelenburg capable No for 3 function Yes for 5 function Mattress length 80 mattress deck length 80 203 cm Mattress length 84 mattress deck length 84 213 cm Mattress length 88 ma...

Страница 14: ...rd in it 5 Remove ties from pendant Before plugging the pendant into the bed frame cut the black O ring at the end of the pendant so it will snap into the pendant housing and fit securely into the bed...

Страница 15: ...ainers four narrow retainers for the sides of the bed and two wider retainers for the end of the bed Insert the two end mattress retainers in the holes appropriate for the mattress length being used L...

Страница 16: ...embly at both ends of the bed frame Place the mounting brackets of the headboard footboard into the slots of the support assembly Slide headboard footboard down until it stops Screw the knobs back in...

Страница 17: ...ment with rails may cause injury or death To help prevent entrapment you should follow the FDA s Entrapment Guidance and your healthcare provider s recommendations when choosing a bed mattress and acc...

Страница 18: ...th assist devices may cause injury or death To help prevent entrapment you should follow the FDA s Entrapment Guidance and your healthcare provider s recommendation to choose a bed mattress and access...

Страница 19: ...The width extension expands by using the push button mechanism Push the expansion button mechanism in and pull out width extension until you reach the desired width of the deck 39 Width Setting 42 Wi...

Страница 20: ...on the crossbar under the head of the bed The Wall Guard should now be in place Place the Wall Guard into the holes of the mount underneath the head section of the bed frame Place pins on each side of...

Страница 21: ...cal personnel 5 Function Staff Control Pad Operation available only on 5 function bed This controller is located on the footboard 1st Set of Buttons Head deck up and down 2nd Set of Buttons Foot deck...

Страница 22: ...leave a bed unattended while the caster locks are not engaged Caster Lock To lock the bed frame push down on all four caster locks with your foot CAUTION Moving the bed while the floor lock or caster...

Страница 23: ...bed Battery is not to be used in a damp environment such as outdoors WARNING The battery backup should be used for EMERGENCY PURPOSES ONLY The only compatible battery backup is a DewertAG7 battery DO...

Страница 24: ...t expose the battery to heat or open flames Avoid storing it in direct sunlight If the battery leaks and you come into contact with the fluid wash the affected area thoroughly in water and consult a...

Страница 25: ...to make sure actuators do not bind at any point throughout their full range of motion Authorized Accessories Inspect all fasteners for looseness and wear Replace or tighten as necessary Ensure welds d...

Страница 26: ...nplug power supply cords Ensure all electrical parts motors control boxes wires and pendant are not broken Ensure NO liquids enter electrical components Remove all gross solid contaminants then wash a...

Страница 27: ...bed Replace tie wraps to hold power cord to frame Plug mains cable into wall outlet Test function WARNING When routing cables in the medical bed precautions shall be taken to avoid squeezing between...

Страница 28: ...mide Solvent Red Pendant Housing PA 765A Rail Release Handle ABS Headboard Footboard Wood Trapeze PVC Steel Nylon Webbing General information The recommended environment for operation and storage tran...

Страница 29: ...d wastes throughout the EU To prevent possible harm to the environment or human health from uncontrolled waste disposal recycle it responsibly to promote the sustainable reuse of material resources To...

Страница 30: ...bed doesn t work continue troubleshooting Blown fuse in control box The control box features time delay self resetting fuses that do not need replacement To reset the fuse unplug the bed for 5 minute...

Страница 31: ...hat the bed may get caught on Reposition the bed or move object causing the obstruction Lower the bed all the way down until all movement has stopped The bed should now be level and remain level when...

Страница 32: ...es Harmonic Emissions IEC 61000 3 2 Not Applicable Voltage fluctuations flicker emissions IEC 6100 3 3 Recommended separation distances between portable and mobile RF communications equipment and the...

Страница 33: ...fixed RF transmitters a as determined by an electromagnetic site survey should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked w...

Страница 34: ...od identified below whichever is earlier Warranty Period Parts Only Anticipated Usable Device Life Frame 10 years on frame 20 years on welds 5 years Electrical Parts Components 1 year 1 year Anticipat...

Страница 35: ...uality Panacea 6300 Bariatric Bed and enjoy outrageous customer service If you have any questions about the bed you have purchased or would like to request warranty service please contact Direct Suppl...

Страница 36: ...rect Supply Inc Distributed by Direct Supply Manufacturing Inc 6699 N Industrial Road Milwaukee WI 53223 2010 2015 Direct Supply Manufacturing Inc All Rights Reserved Direct Supply Panacea and all ass...

Отзывы: