Servicing
DANGER SHOCK HAZARD
Possible shock hazard may occur if the main power supply cord for the control box is not unplugged from
the wall outlet before any maintenance is performed on electrical components.
Actuators and Control Box
Cord and Socket Identification
Replacing Control Box
• Unplug mains cable from the wall outlet
• Unplug connections for all actuators
and accessories
• Separate the control box from the actuator
• Reinstall the control box on the actuator
• Plug all connections back into control box
following color coding
• Plug mains cable into wall outlet
• Test function
WARNING
: When routing cables in the medical
bed, precautions shall be taken to avoid squeezing
between parts of the medical bed.
Replacing the AC Power Cord (mains cable)
• Unplug mains cable from wall outlet
• Unplug mains cable from the control box
• Remove tie-wraps holding power cord in place
on frame
• Plug the new power cord into the control box
• Route the new power cord in the same manner
as the original cord, ensuring use of strain relief
device at head end of bed
• Replace tie-wraps to hold power cord to frame
• Plug mains cable into wall outlet
• Test function
WARNING
: When routing cables in the medical
bed, precautions shall be taken to avoid squeezing
between parts of the medical bed.
Replacing Actuators
• Unplug mains cable from the wall outlet
• Identify the actuator to replace
• With the help of a second person, place bed on
its side to remove the actuators
• Unplug actuator cord from control box
• Actuator is held in place by (2) clevis pins
• Remove bowtie clips from clevis pins
• Slide the clevis pins out of the holes
• Replace the actuator
• To reassemble bed, reverse previous steps, and
make sure to:
- Assemble clevis pins as originally installed with
bowtie clips
- Zip ties should be replaced, with cords in their
original position and routing direction to the
control box
• Plug mains cable into wall outlet
• Test function
DANGER
: Failure to replace bowtie clips on
clevis pins at each end of actuator can result in
harm to the user, operator and/or equipment.
WARNING
: When routing cables in the medical
bed, precautions shall be taken to avoid squeezing
between parts of the medical bed.
Bed End Panel
(Green)
Hand Pendant
(Red)
Head Section Motor
(Black)
Foot Section Motor
(Yellow)
Foot End Hi-Lo Motor
(Blue)
Head End Hi-Lo Motor
(White)
Battery Backup
(Black)