7. Control Box Wiring

Remove the control box cover to make the connections as described 
below. Use a screwdriver or similar; refer to the special labels on the 
control box. 

NOTE! Make sure that plugs are connected to the correct sockets 
otherwise the actuators can be damaged.

1.  Battery terminal socket.

2.  Plug socket. NOTE: The plug must ALWAYS be connected to the 

control box for IP protection to be valid.

3.  Remote control socket.

4.  Socket for the raise/lower actuator. 

8. Battery Charging

Battery charging must take place when the Walker is not in use (the battery 

can never be overcharged). Charging is carried out by inserting the plug in 
a wall socket.

The battery should be charged at least 24 hours before first time use.

9. LED on Control Box and Battery

Control box – green light when power is available from battery/mains 

socket.

Battery:

•  The LED flashes green during charging.

•  ON/OFF: To use the product, press ON.

•  The green lamp will begin to flash.

•  Use the OFF button (hold down for 5 s) to switch off the battery if the 

product will not be used for an extended period.

•  The LED will shows a constant green light when the battery is fully 

charged.

•  When the battery is in normal state/in operation, the LED lashes green 

at REGULAR intervals, namely 0.5 s ON, 4 s OFF.

•  When the battery requires charging (remaining charge <25%) the LED 

will flash orange/red at REGULAR intervals, namely 0.5 s ON, 4 s OFF.

NOTE: The battery has an in-built device that automatically switches it off 
when remaining charge is <20%. The LED is then extinguished. The battery 
must now be fully charged before the electronics can be used again.

1

2

3

4

Содержание 56-315

Страница 1: ...nual DIRECTHEALTH C ARE GR OUP COM Manual no 801 rev 1 210604 SWL 240 kg 529 lbs 56 315 56 315 75 56 315 75 UK 56 315 75 PU 56 315 PU AUS 56 315 AUS 56 315 CAN 56 315 PU 56 315 PU UK 56 315 B 0 56 315 B 1 56 315 B 0 PU 56 315 S 56 315 UK 56 315 100 ...

Страница 2: ...5 10 Maintenance 06 11 General Maintenance Advice Cleaning 07 12 Technical Specifications 07 13 Safety Instructions 08 14 Recycling Disposal 08 15 Warranty 08 16 Electrical System Technical Data T motion Linak 09 17 Parts List 10 18 Spare Parts 11 19 Accessories 11 20 Table of Symbols 12 Bure XL Bure XL was designed for the bigger heavier patient and therefore provides extra space and strength Bec...

Страница 3: ... is not in use Check the function of the brakes before use Inspect the device regularly to detect any signs of damage If there are any visible signs of damage contact Direct Healthcare Group Customer Service In case of an electrical model of the device check that the hand control does not show signs of wear before every use Check that the hand control markings are in accordance with the lifting fu...

Страница 4: ...e Group representative When lifting the device only use the denoted handholds At least two people are required for lifting the device Do not leave a user or patient unattended when the device is being used Never move the lift by pulling on the actuator or gas piston The device must not be lowered into water The device must not be left or stored in a damp or humid environment The device must not be...

Страница 5: ...ast 24 hours before first time use 9 LED on Control Box and Battery Control box green light when power is available from battery mains socket Battery The LED flashes green during charging ON OFF To use the product press ON The green lamp will begin to flash Use the OFF button hold down for 5 s to switch off the battery if the product will not be used for an extended period The LED will shows a con...

Страница 6: ...rom play Height adjustment via remote control is functional The electric actuator must run smoothly at a constant speed Fit new guide sleeves in frame or end plugs in chrome tubes as required Tighten all lock bolts Replace clamps Tighten the attachment fitting concerned replace bolts lock nuts as necessary Check that the battery remote control actuator are connected in accordance with the wiring d...

Страница 7: ...ifetime Replace damaged parts immediately must be done by experienced personnel If the product is to be cleaned in a washing tunnel washing cabin use PU cushions The product is only intended for indoor use 12 Technical Specifications Dimensions apply to Walkers with 125 mm castors Part No Product Product width Max user weight Length Height Armrest Internal dimension 56 315 Bure XL EL 90cm 240kg 84...

Страница 8: ...ncluding batteries must be removed and recycled as electrical components Ask your local recycling station for further information about how the different materials in the product metals plastics electronics should be recycled 15 Warranty We provide two years warranty against manufacturing faults does not apply to wear parts For further information or enquiries please contact Direct Healthcare Grou...

Страница 9: ...00N 1200 N Self locking push 1000N 1000N 7A 3 A Current 6 6 1 0 A 2 6 0 4 A 7A ut 7 A in 3A ut 3 A in Colour Black Aluminium White White Stroke 400mm 55mm 400 mm 55 mm Mounting dim 605mm 167mm 606 2 mm 165 2 mm Encapsulation IP66 IP66 IPX6 IPX6 IEC60601 1 3rd revision Control box Battery Producer T Motion Linak Voltage 24V 230V 24V Capacity 2 0AH 2 25 Ah Encapsulation IP66 IPX6 Colour Black White ...

Страница 10: ...t Washer 2 D E F Handle and grip Clamp Knob 7 Protective plate left right 3 Upper frame 8 S T U V Remote control Holder remote control Battery Control box 4 G H I J Chromium plated tube Attachment bolt Washer End bushing 9 X Y Actuator raise lower Bolt washer 2 pcs nut attachment at bottom and top 5 K L Frame with battery shelf Guide sleeve ...

Страница 11: ...6 357 Wheel without brake 100 mm 56 358 Wheel with brake 100 mm N 56 360 Wheel without brake 125 mm 56 361 Wheel with brake 125 mm 56 387 Wheel with brake 75 mm E 56 356 Locking clamp handle F 56 363 Knob for adjusting handle B 56 364 Knob for adj width of armrest plates S 56 370 1 Hand control 2channel raise lower LA31 S 56 370 Hand control 2channel raise lower LA28 56 374 1 Power cable LA31 56 3...

Страница 12: ...Medical Device Read the manual Caution Product Code Manufacturer information Batch Code Type BF according to the degree of protection against electric shock The device is intended for indoor use WEEE Symbol May not be discarded in domestic waste Class II Equipment Duty Cycle 2 min in active ON mode 18 min in rest OFF mode 20 Table of Symbols ...

Страница 13: ...atan 35 SE 164 40 Kista Sweden Tel 46 0 8 557 62 200 info directhealthcaregroup com www directhealthcaregroup com Medical Device Class I The product complies with the re quirements of the Medical Devices Regulation 2017 745 Medical Device Class I ...

Отзывы: