4.5 Peak Ghosting
Ghosting is the appearance of extraneous peaks in a
chromatogram. These may be late-eluting peaks from a previous
injection or result from a contaminated valve or a poor sample
loading technique. These peaks may co-elute with peaks of
interest, resulting in non-reproducible peak heights.
•
Insufficient time between sample injections.
Wait until the previous sample has been completely eluted
before making another injection.
•
Insufficient flush between samples.
Flush the sample loop with at least 10 loop volumes of
deionized water or sample between sample injections.
4.6 Non-Reproducible Peak Height or Retention Time
•
Column overloading.
1. Change to a sample loop with a smaller volume.
2. Dilute the sample.
•
Liquid leaks.
Locate and eliminate the leaks.
Also see “Peak ghosting” in Section 4.5 above.
4.7 Abnormal Retention Time or Selectivity
•
System is not equilibrated following an eluent change.
Allow the system to equilibrate with at least 20 column
volumes of eluent (for example, 30 minutes at 2.0 mL/min
for 4 mm anion separator columns).
•
Flow rate through system is incorrect.
1. Check that the correct flow rate is selected.
LC20 Chromatography Enclosure
4-6
Doc. 034859-02 9/93
Содержание LC20
Страница 6: ...Contents iv Doc 034859 02 9 93 ...
Страница 8: ...LC20 Chromatography Enclosure 1 2 Doc 034859 02 9 93 ...
Страница 12: ...LC20 Chromatography Enclosure 2 2 Doc 034859 02 9 93 ...
Страница 24: ...LC20 Chromatography Enclosure 2 14 Doc 034859 02 9 93 ...
Страница 26: ...LC20 Chromatography Enclosure 3 2 Doc 034859 02 9 93 ...
Страница 29: ...Figure 3 7 Drawing Sample 3 Operation and Maintenance Doc 034859 02 9 93 3 5 ...
Страница 32: ...LC20 Chromatography Enclosure 3 8 Doc 034859 02 9 93 ...
Страница 34: ...LC20 Chromatography Enclosure 4 2 Doc 034859 02 9 93 ...
Страница 40: ...LC20 Chromatography Enclosure 4 8 Doc 034859 02 9 93 ...
Страница 42: ...LC20 Chromatography Enclosure 5 2 Doc 034859 02 9 93 ...
Страница 50: ...LC20 Chromatography Enclosure 5 10 Doc 034859 02 9 93 ...
Страница 52: ...LC20 Chromatography Enclosure A 2 Doc 034859 02 9 93 ...
Страница 56: ...LC20 Chromatography Enclosure B 2 Doc 034859 02 9 93 ...
Страница 65: ...Figure B 6 Electrochemical Cell Plumbing Schematic B Installation Doc 034859 02 9 93 B 11 ...