3
•
Operation and Maintenance
3.1 Routine Operation
Inside the Rheodyne injection valve, eluent flows through one of
two paths depending on the position of the injection valve knob (see
Figure 3-6).
In the
LOAD
position, eluent flows in from the pump and out to the
column without entering the sample loop. Sample flows from the
syringe into the valve and through the sample loop, and excess
sample flows out to waste.
In the
INJECT
position, eluent flows in from the pump, through the
sample loop, and out to the column (carrying the contents of the
sample loop with it).
Figure 3-6. Injection Valve Schematic
Doc. 034859-02 9/93
3-3
Содержание LC20
Страница 6: ...Contents iv Doc 034859 02 9 93 ...
Страница 8: ...LC20 Chromatography Enclosure 1 2 Doc 034859 02 9 93 ...
Страница 12: ...LC20 Chromatography Enclosure 2 2 Doc 034859 02 9 93 ...
Страница 24: ...LC20 Chromatography Enclosure 2 14 Doc 034859 02 9 93 ...
Страница 26: ...LC20 Chromatography Enclosure 3 2 Doc 034859 02 9 93 ...
Страница 29: ...Figure 3 7 Drawing Sample 3 Operation and Maintenance Doc 034859 02 9 93 3 5 ...
Страница 32: ...LC20 Chromatography Enclosure 3 8 Doc 034859 02 9 93 ...
Страница 34: ...LC20 Chromatography Enclosure 4 2 Doc 034859 02 9 93 ...
Страница 40: ...LC20 Chromatography Enclosure 4 8 Doc 034859 02 9 93 ...
Страница 42: ...LC20 Chromatography Enclosure 5 2 Doc 034859 02 9 93 ...
Страница 50: ...LC20 Chromatography Enclosure 5 10 Doc 034859 02 9 93 ...
Страница 52: ...LC20 Chromatography Enclosure A 2 Doc 034859 02 9 93 ...
Страница 56: ...LC20 Chromatography Enclosure B 2 Doc 034859 02 9 93 ...
Страница 65: ...Figure B 6 Electrochemical Cell Plumbing Schematic B Installation Doc 034859 02 9 93 B 11 ...