ICS-2100 Installation Instructions
30
Doc. 065294-01 3/09
13.4
Installing the Suppressor
1. Before installing the suppressor, hydrate it as instructed in the
suppressor manual. Suppressor manuals are provided on the Dionex
Reference Library CD-ROM (P/N 053891).
2. Locate the suppressor
mounting tabs on the
component panel (see
). Reach into
the opening between the
tabs and pull the yellow
connector and attached
cable slightly out of the
opening.
Connect this cable to the
cable from the
suppressor (align the
pins on the two
connectors and push
them together).
Figure 29. Suppressor Mounting Tabs
and Guidelines
3. Orient the suppressor with the
REGEN OUT
fitting on top and the
cable to the right.
4. Feed the cable into the opening in the component panel.
5. Align the top of the suppressor with the guideline printed on the
component panel (see
) and align the slots on the rear of the
suppressor with the tabs on the panel.
1
1
2
3
4
1
Suppressor
Mounting Tabs
Suppressor
Cable Connector
Guideline for
AES
®
Guideline for
SRS
®
1
2
3
4