Connecting the Waste Lines
Doc. 065294-01 3/09
31
6. Press the suppressor onto the tabs and
then slide it down to lock it in place. Pull
slightly on the center of the suppressor to
verify that it is securely fastened.
®
suppressor
installed on the component panel.
7. Remove the coupler connecting the
ELUENT IN
and
ELUENT OUT
lines and
connect these lines to the corresponding
ports on the suppressor.
Figure 30. Suppressor Installed
(SRS Shown)
8. Remove the fitting plug from the
REGEN IN
line and connect it to the
REGEN IN
port on the suppressor.
9.
Do not connect the
REGEN OUT
port until after completing the
eluent generator setup (
.
10. Replace the cover on the column heater, making sure the
GUARD IN
line is aligned with the slot on the side of the heater cover.
14.
Connecting the Waste Lines
1. Untape the coiled waste lines from the rear panel. Place the ends of the
PUMP
WASTE
and
TO WASTE OUT
lines into a waste container.
2. Direct the
WASTE, GAS SEPARATOR
line from the top of the IC system to the
rear panel. Snap the line onto one of the tubing clips on the rear panel.
Connect the line to the waste, gas separator tube (see
NOTE To prevent waste from siphoning back into the system,
check the lines periodically to be sure they are not bent,
pinched, or elevated at any point.