ASE 150 Operator’s Manual
62
Doc. 065207-02 9/08
Error 013
Clamp plate temperature high.
Cause:
The heater cable connection to the power supply is loose, or a heater
component malfunctioned.
Action:
Contact Dionex for assistance.
Error 014
Oven unable to reach temperature.
Cause:
Incorrect settings of the oven voltage switches on the ASE 150 rear
panel.
Action:
Set both oven voltage switches to match the voltage from the power
source at the installation site: either
110
(for voltages between 100
and 120 Vac) or
220
(for voltages between 220 and 240 Vac). For
details, see
.
Cause:
The heater cable connection to the power supply is loose, or a heater
component malfunctioned.
Action:
Contact Dionex for assistance.
Error 015
Unable to rinse.
Cause:
The pump could not deliver the specified rinse volume within the
allotted time. The cell or lines may be plugged.
Action:
Run a rinse cycle (see
). If the rinse runs without error, the
sample cell may be plugged. Replace the cell filter (see
Action:
If the rinse has the same error, there may be a blockage in the solvent
lines. Contact Dionex for assistance.