ASE 150 Operator’s Manual
110
Doc. 065207-02 9/08
B.2.2
Connecting the Drain Hose
1. Push one end of the convoluted hose (P/N 055075) onto the drain
fitting on the ASE 150 rear panel.
2. Route the free end of the hose to waste.
B.2.3
Checking the Oven Voltage Switches
1. Verify that the settings of both oven voltage switches on the ASE 150
) match the voltage from the power source
at the installation site: either
110
(for voltages between 100 and
120 Vac) or
220
(for voltages between 220 and 240 Vac).
2. If you need to change an oven voltage switch setting:
a. Verify that the main power switch on the ASE 150 rear panel is
turned off.
b. Insert a small screwdriver into the slot in each switch and turn it
to either
110
or
220
.
NOTE The oven voltage switch settings affect only the
oven and have no effect on any other system
components. The ASE 150 main power supply
adjusts automatically to the power source at the
installation site.
Figure B-2. Oven Voltage Switches
Oven Voltage
Switches