ASE 150 Operator’s Manual
58
Doc. 065207-02 9/08
Error 002
System pressure low.
Cause:
Insufficient nitrogen gas pressure applied to the system.
Action:
Check the
REGULATORS
SYSTEM
field should read 0.34
±
0.02 MPa (50
±
3 psi). If the
SYSTEM
field
is low, check the gas pressure supplied to the ASE 150. It should be
between 0.97 and 1.38 MPa (140 and 200 psi); 1.03 MPa (150 psi) is
recommended. If necessary, replace the gas cylinder.
Action:
If the error message reappears, contact Dionex for assistance.
Error 003
Oven temperature low.
Cause:
The heater cable connection to the power supply is loose, or a heater
component malfunctioned.
Action:
Contact Dionex for assistance.
Error 004
Oven temperature high
Cause:
The heater cable connection to the power supply is loose, or a heater
component malfunctioned.
Action:
Contact Dionex for assistance.
Error 005
Cell pressure threshold exceeded.
Cause:
A blockage in the system.
Action:
Run a rinse cycle (see
). If the rinse runs without error, the
sample cell may be plugged. Replace the cell filter (see
). If the error still appears, clean or replace the cell frit
Action:
If the rinse has the same error, there may be a blockage in the solvent
lines. Contact Dionex for assistance.
Содержание ASE 150
Страница 1: ...ASE 150 Accelerated Solvent Extractor Operator s Manual Document No 065207 Revision 02 September 2008...
Страница 32: ...ASE 150 Operator s Manual 24 Doc 065207 02 9 08...
Страница 64: ...ASE 150 Operator s Manual 56 Doc 065207 02 9 08...
Страница 76: ...ASE 150 Operator s Manual 68 Doc 065207 02 9 08...
Страница 114: ...ASE 150 Operator s Manual 106 Doc 065207 02 9 08...
Страница 132: ...ASE 150 Operator s Manual 124 Doc 065207 02 9 08...