B • Installation
B.1 Facility Requirements
•
Install the ASE 150 Accelerated Solvent Extractor on a sturdy table or
workbench in a location that minimizes ASE 150 exposure to direct sunlight.
Allow at least 6 cm (2.5 in) of free space behind the ASE 150 for connections
and ventilation.
•
Provide a source of 99.9% pure nitrogen gas, regulated to between 0.97 and
1.38 MPa (140 and 200 psi); 1.03 MPa (150 psi) is recommended.
Applications that use a very clean baseline electron capture detector (ECD)
may require UHP (ultra-high purity) gas.
•
Provide a grounded, single-phase power source of 100 to 240 Vac, 50/60 Hz.
The oven voltage switches must be set to match the voltage from the power
source at the installation site:
110
(for voltages between 100 and 120 Vac) or
220
(for voltages between 220 and 240 Vac). For details, see
.
Lift the ASE 150 only from the bottom and/or sides of the instrument.
If the system is lifted from the front, the drip tray may slide forward.
Use caution when lifting the ASE 150: it weighs 34 kg (75 lb).
Ne soulevez le ASE 150 que par le fond ou les côtés. Soyez prudent
lorsque vous soulevez le ASE 150: il pèse 34 kg.
Wenn Sie den ASE 150 anheben oder bewegen möchten, greifen Sie
bitte unter den Boden oder heben Sie das Gerät an den Seiten an.
Seien Sie vorsichtig, wenn Sie den ASE 150 anheben. Das Gerät wiegt
34 kg.
Содержание ASE 150
Страница 1: ...ASE 150 Accelerated Solvent Extractor Operator s Manual Document No 065207 Revision 02 September 2008...
Страница 32: ...ASE 150 Operator s Manual 24 Doc 065207 02 9 08...
Страница 64: ...ASE 150 Operator s Manual 56 Doc 065207 02 9 08...
Страница 76: ...ASE 150 Operator s Manual 68 Doc 065207 02 9 08...
Страница 114: ...ASE 150 Operator s Manual 106 Doc 065207 02 9 08...
Страница 132: ...ASE 150 Operator s Manual 124 Doc 065207 02 9 08...