5 • Service
Doc. 065207-02 9/08
99
5.9
Replacing the Source Needle
1. Toggle the needle switch to the
DOWN
position.
2. Follow the instructions in
to disconnect the gas source and
remove the right-side panel.
3. Use a 2.5-mm hex wrench to
remove the two screws on the
top of the needle block (see
Figure 5-22. Removing the Needle Block Screws
4. Slide the cover of the needle
block to the front and right,
away from the needle (see
cover.
Figure 5-23. Removing the Needle Block Cover
Remove
these
screws
Needle
Block
Needle
Block Cover