7. Disconnect the red tubing and the transparent blue tubing from the
two black elbow press fittings on top of the pump (see Figure 5-6).
8. Using a 1/4-inch open-end wrench, disconnect the stainless steel
fitting from the outlet check valve (see Figure 5-6).
9. Disconnect the PEEK fitting from the inlet check valve, by hand (see
10. Disconnect the gray cable from the pump PC board (see Figure 5-6).
11. Using a 3 mm hex-screw driver (P/N 060154), remove the screws in
the left and right end plates (see Figure 5-6) that secure the pump to
the component panel.
Figure 5-6. ASE 100 Pump