Understanding the Status Display During a Run
During runs, the
MAIN STATUS
and
DETAIL STATUS
screens display the
operations currently being performed. Because of sample prep overlap,
information on multiple injections may be displayed simultaneously at
certain points in the schedule. For example, in
, the AS is
injecting the sample and performing the timed events for the first
injection from vial 1 while preparing for the second injection.
Notes
•
The
INJ#
field is updated when the AS loads the loop for the next
injection.
•
The
TIME
field and the
METHOD
timed events clock are reset to zero at
injection.
•
The
TIME
field counts up from the injection until the next injection
occurs.
•
The
METHOD
timed events clock counts up until all timed events have
been performed for an injection. The clock then stops and is no longer
displayed.
Figure 2-21. Detail Status Screen Example
Help Prompt
FLUSHING
VIAL#:
TIME:
INJ#:
VOL:
METHOD:
TRAY:
LOOP:
12.5
12.53
25
100
1
1/3
1
2 mL
uL
uL
min
TTL1
TTL2
RLY1
RLY2
0
1
0
0
L
NORMAL PARTIAL
SCH 1
LOCAL
DETAIL
LINE 10
TRAY: 25 20 °C
INJ VLV
Preparing for
Injection #2
Elapsed Time
Since Injection #1
Method Timed Events
Clock for Injection #1
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...