![Dini Argeo MCW PROFESSIONAL Скачать руководство пользователя страница 18](http://html1.mh-extra.com/html/dini-argeo/mcw-professional/mcw-professional_instruction-manual_2505724018.webp)
MCW-MCWR2-MCWK PROFESSIONAL MCW-MCWK_03_12.10_EN_U
18
1.2.4 Electronic device features
PROTECTION DEGREE IP 67
POWER SUPPLY
6 Vdc – 4,5 Ah rechargeable internal battery, standard, range of
about 60 hours. 12 Vdc with external power supply 100 ÷ 240 Vac
(50÷60 Hz)/12 Vdc standard.
MAXIMUM POWER
5 VA.
MINIMUM VOLTAGE PER DIVISION
0.3
µ
V (Homologated instrument); 0.03
µ
V (non homologated
instrument).
DISPLAYED DIVISIONS
10000e, 3X3000e for legal for trade use expandable to 800.000
for internal factory use (with
minimum
signal coming from the
1,6mV/V cell).
DISPLAY
MCW - Backlit 25 mm LCD display with 6 digits.
MCWK - 40 mm LED display with 6 digits.
RESOLUTION IN COMTAGE
1'500’000 points (with input signal equal to 3mV/V).
KEYBOARD
Water resistant key polycarbonate membrane with tactile
feedback.
TARE FUNCTION
Subtractive on the entire capacity.
AUTO POWER OFF
Adjustable from 1 to 255 minutes of no use, disinsertable.
LOW BATTERY WARNING
"Low bat" shown on the display.
MEASUREMENT UNIT AVAILABLE
g= grams, kg= kilograms, t= tons, Lb= pounds.
IN CALIBRATION PHASE
CONTAINER
Sturdy stainless steel, against dust and splashes.
POWER OF LOAD CELLS
5Vdc ± 5%, 120mA (max 8 load cells 350 Ohm).
I/O SECTION
- 1 RS232/TTL input/output
- 1 RS232/input/output
Configurable for connection to PC/PLC, WEIGHT REPEATER or.
PRINTER.
THEORETICAL LIFE
If the instrument is regularly maintained and if the user instructions
shown in this manual are carried out, the instrument will attain a
theoretical life of 5 years; the data is a function of use
is subject to change according to audits by the manufacturer.
For further details, see section “MAINTENANCE AND
VERIFICATIONS”.
THE PARTS OF THE INSTRUMENT CONTAINING DANGEROUS ELECTRICAL TENSION ARE ISOLATED AND
INACCESSIBLE TO THE USER UNLESS IT HAS BEEN DAMAGED, OPENED, OR ALTERED.