Installation

Installation requirements

Observe the following conditions for correct installation of the industrial PC and of the load receiver: 

• 

Flat, level support surface.

• 

Stability and absence of vibrations.

• 

Absence of aggressive dust and vapours.

• 

Absence of draughts.

• 

Make sure that the platform is levelled or that the load cells are evenly supported.

• 

Moderate temperature and humidity (15°C - 30°C and 40% - 70%).

• 

Do not install in an environment where there is a risk of explosion.

• 

All transmitter connections must be made in accordance with applicable regulations in the area and environment of installation. 

Observe the electrical precautions listed in the section “

Electrical precautions

”.

• 

Ensure that it is correctly earthed, see the relevant section “

Earthing of the system

”.

• 

Do not perform welding when the load cells have already been installed.

• 

If necessary, use watertight sheaths and fittings to protect the load cell cables.

• 

Any junction boxes must be watertight.

• 

Anything not expressly described in this manual constitutes improper use of the equipment.

Electrical precautions

• 

Use a regulated mains supply within ± 10% of the rated voltage.

• 

The electrical protections 

(fuses, etc.)

 are the responsibility of the installer.

• 

Observe the recommended minimum distances between cables of different categories 

(see table on page 10)

.

• 

The following cables must comply with the maximum permissible lengths 

(see table on page 10)

, they must be shielded and must 

be inserted alone in metal conduits or pipes: 

- the load cell extension cables; 

- the signal amplifier cables. 

The input of the cell and amplifier cables into the PC case must be independent. The cables must be connected directly to the 

transmitter terminal block.

• 

Fit the “RC” filters on all devices that produce electrical noise. 

• 

Connections to load cells and any external device must be as short as possible.

• 

The cable ends (connectors, leads, terminals, etc.) must be installed correctly; the cable shielding must be kept intact until close to 

the connection point.

The sticker shown by the side is applied to the product to warn the user of the electrical hazard:

• 

- RISK OF ELECTRIC SHOCK

• 

- DO NOT DISASSEMBLE THE APPLIANCE

• 

- COMPLETELY DISCONNECT THE POWER SUPPLY TO THE APPLIANCE BEFORE PERFORMING ANY MAINTENANCE.

Содержание 6116EVO

Страница 1: ...www diniargeo com 6116EVO Industrial touchscreen PC with integrated scale input ENGLISH USER MANUAL...

Страница 2: ......

Страница 3: ...lation 5 Installation requirements 5 Electrical precautions 5 Earthing of the system 7 Dimensional technical drawings 8 6116EVO 6116EVO PC 8 6116EVO DGT 9 Wiring diagrams 10 6116EVO 10 6116EVO DGT 11...

Страница 4: ...y the greatest attention to all those paragraphs that illustrate the simplest and safest way to operate The utmost care has been taken in compiling this manual but reports of any inaccuracies are alwa...

Страница 5: ...d in this manual constitutes improper use of the equipment Electrical precautions Use a regulated mains supply within 10 of the rated voltage The electrical protections fuses etc are the responsibilit...

Страница 6: ...tres with 2 x 0 5 mm2 cable 150 metres with 2 x 1 mm2 cable RECOMMENDED DISTANCES AND CABLE CLASSIFICATION MAXIMUM ALLOWED LENGTHS Category I Category II Category III Category IV Distance 100 mm 200 m...

Страница 7: ...t release material onto the weighing structure using cables with a cross section of not less than 16 mm2 Also connect the upper part with the lower part of each cell by means of a copper braid with a...

Страница 8: ...8 Optimized layout for A4 print 329 100 100 66 233 211 327 379 USB ETHERNET VESA 100 M4 x 10 mm PG9 2 7 26 140 7 11 151 354 6116EVO 6116EVO PC USER_MAN_ENG_6116EVO Dimensional technical drawings...

Страница 9: ...9 Optimized layout for A4 print 329 100 100 120 120 66 15 5 233 211 327 379 USB ETHERNET PG9 151 354 2 7 26 140 7 11 6116EVO DGT USER_MAN_ENG_6116EVO...

Страница 10: ...timized layout for A4 print RS485 Cell 4 Cell 3 Cell 2 Cell 1 Power 485 POWER CELL ETHERNET USB 2 0 USB 2 0 POWER CELL 3 CELL 2 ETHERNET CELL 1 USB 2 0 USB 2 0 6116EVO USER_MAN_ENG_6116EVO Wiring diag...

Страница 11: ...11 Optimized layout for A4 print POWER CELL ETHERNET USB 2 0 USB 2 0 6116EVO DGT USER_MAN_ENG_6116EVO...

Страница 12: ...SUPPLY CELL 1 CELL 4 CELL 3 CELL 2 B Vdc GND Vdc GND EXC EXC Sen Sen Sig1 Sig1 Exc Exc Sig2 Sig2 Exc Exc Sig3 Sig3 Exc Exc Sig4 Sig4 K1 K2 LOAD CELL 1 E E S S E E S S SIG SIG SEN SEN EXC EXC LOAD CELL...

Страница 13: ...AL 4 Sig4 LOAD CELL E E S S SIG SIG SEN SEN EXC EXC LOAD CELL 2 4 SIG 12 24 Vdc 18 17 16 6 7 8 3 4 5 1 2 13 12 15 14 11 10 LOAD CELL RS232 RS485 POWER B A 9 GND RX TX DISPLAY J1 CAL J1 M DGT1 USER_MAN...

Страница 14: ...ntrol WORKING TEMPERATURE 0 60 C 0 40 C for type approved applications 6116EVO DGT only DIMENSIONS Refer to the dimensional technical drawings POWER AND CONSUMPTION 110 240 Vac max 75 W OIML SCALE INP...

Страница 15: ...Sig1 Sig1 Exc Exc K1 SW1 CELL 3 Exc Sig4 Sig4 LD1 SW1 1 CELL 3 CELL 3 CELL 2 Sig1 Sig1 Exc Exc Sig2 Sig2 Exc Exc Sig3 Sig3 Exc Exc Sig4 Sig4 A B A TERM RS485 PWR SUPPLY CELL 1 CELL 2 B Vdc GND Vdc GND...

Страница 16: ...FIELDBUS 25 GND 16 15 14 13 12 11 CELL SIG EXC EXC SIG SIG SEN SEN 10 USER_MAN_ENG_6116EVO 6116EVO DGT The transmitter has a channel CELL for connection to 6 wire load cells with the use of SENSE The...

Страница 17: ...le takes place through ASCII serial protocol or Modbus RTU Check the serial port to be used for communication using the Windows Device Manager program in the Ports COM LPT submenu For more information...

Страница 18: ...ves the right to make changes to its products at any time without notice and without incurring any penalty We therefore recommend that you always check for updates The person responsible for the use o...

Страница 19: ......

Страница 20: ...536 921784 Fax 39 0536 926654 www diniargeo com The information in this document is approximate and can be subject to variations without prior notice by Dini Argeo in compliance with the norms in forc...

Отзывы: