3590ET - 3590EGT_Serial_Commands
37
PID
Description
Stores weigh data in the alibi memory and get alibi ID value.
Format
P
I
D
Example
Data stored in alibi with a gross weight equal to 15 kg and a preset tare of 1 kg when the instrument is on
channel 1.
Command
P
I
D
Answer
P
I
D
S
T
,
1
,
1
5
.
0
0
0
k
g
,
P
T
1
.
0
0
0
k
g
,
0
0
0
0
0
-
0
0
0
0
0
5
Answer
P
I
D
s
s
,
c
,
w
w
w
w
w
w
w
w
w
w
u
u
,
p
p
t
t
t
t
t
t
t
t
t
t
u
u
,
r
r
r
r
r
-
n
n
n
n
n
n
Where
ss
E
R
Remote scale selected and remote scale disconnected.
T
L
Tilt condition error.
O
L
Over load condition.
U
L
Under load condition.
S
T
Weight stable.
U
S
Weight unstable.
c
Instrument channel.
w...w
Gross weight on 10 characters padded with blank spaces on front.
uu
Unit of measure (“g”, “kg”, “t”, “lb”).
pp
Tare type (“ ” with semi-automatic tare, “PT” with preset tare).
t...t
Tare value.
r…r
Alibi rewrite ID value on 5 digits padded with zeroes on front.
n…n
Alibi ID value on 6 digits padded with zeroes on front.
In case of error with no weight data stored in alibi memory in place of rrrrr-nnnnnn there is
N
O
Содержание 3590EGT Series
Страница 1: ...www diniargeo com 3590ET 3590EGT ENGLISH Serial commands TECHNICAL MANUAL ...
Страница 2: ......
Страница 5: ...WUBU 54 X 55 Z 56 ZERO 56 6 Simple example 57 ...
Страница 59: ...3590ET 3590EGT_Serial_Commands 59 ...
Страница 60: ...3590ET 3590EGT_Serial_Commands 60 NOTES ...
Страница 61: ...3590ET 3590EGT_Serial_Commands 61 NOTES ...
Страница 62: ...NOTES ...
Страница 63: ......