15
Lower row
display
Upper row
display
meaning
[
,,,F08
]
[
, 0 3 8 4 0 0
]
2.
Baud
rate
of
communication
port
(60012001800240048009600192003840057600115200
optional)
[
,,,F09
]
[
f lt 0 . 0
]
1. Scale filter coefficient (0-9 optional, the larger
the number, the deeper the filter)
After pressing [input], the filter coefficient of
[flt-2,0] is displayed
After pressing [Enter], it will display [uint, 00.0],
and this parameter is for standby
[
,,,F10
]
[
rt 0 . 2
]
Stability determination time (1.0 s is recommended)
[
,,,F11
]
[
R f 0 1
]
Stable range (1 is recommended)
The larger the value is, the more unstable the weighing
is, such as the livestock scale
[
,,,F12
]
[
c a l- 1
]
Weighing calibration (see Chapter 6 for details)
[
,,,F13
]
[
* * . * * * *
]
Test sensor output signal
Switch the two groups of sensors through
the [→] key
The display [1 * *. * *] indicates the
millivolts of the first group of sensors
The display [2 * *. * *] indicates the
millivolts of the second group of
sensors
See section
8.1 for
details
[
,,,F14
]
[
t e st- d s p
]
Display test
[
,,,F15
]
[
o-000000
]
Test output port
[
,,,F16
]
[
tst- p s u l
]
Test pulse input port
[
,,,F17
]
[
d **.**.**
]
current date
[
,,,F18
]
[
T **.**.**
]
current time
[
,,,F19
]
[
LINE,,**
]
Set workflow number (factory default 00 means none)
1-14 corresponding to the fixed process listed in the
manual
No.15 is an empty process. Note: 1-6 is the fixed
process of instrument and cannot be modified. 7-15 can
receive user written process through serial port.
Please refer to relevant process information for
details
[
,,,F20
]
[
DSP1 ***
]
In weighing state, display the contents on the upper
row of the display panel
Press [Enter] to display [KP1,, * *], indicating the
content of sub display in the lower row
See section 7.5 for details.