7. Then set the voltage of CLM by
trimmer 10 V
onto 10 V (or to the calculated value of U out -
see above).
8. To reach a maximum accuracy is useful to check the setting of 0 V again.
Table of ranges
No.
Sensitivity
2)
(C)
Range
3)
(R)
Position of DIP switch
1
1)
20 pF
70 pF
2
30 pF
250 pF
3
50 pF
600 pF
4
100 pF
1 200 pF
5
150 pF
3 000 pF
6
300 pF
7 000 pF
7
500 pF
18 000 pF
8
1 000 pF
36 000 pF
Tables of recommended ranges depending on the medium and the type of level meter
Medium
Plastic granulates (ε
r
approximately 2)
Type / electrode length
1 m
2 m
5 m
10 m
20 m
10, 30, 31
1
2
2
3
4
Medium
Flour, sand, grains (ε
r
approximately3)
Type / electrode length
1 m
2 m
5 m
10 m
20 m
10, 30, 31
2
2
3
4
5
Medium
Cement (ε
r
approximately 4)
Type / electrode length
1 m
2 m
5 m
10 m
20 m
10, 30, 31
2
2
3
4
5
Medium
Water (water solutions)
Type / electrode length
1 m
2 m
5 m
10 m
20 m
11, 12
3
4
5
x
x
22
3
4
x
x
x
32
3
4
5
6
6
Medium
Benzine, petroleum, diesel, oil
(ε
r
approximately 2)
Type / electrode length
1 m
2 m
5 m
10 m
20 m
10, 30
1
2
2
3
4
20
3
4
x
x
x