Use X-
NIR™ RevA3
ENGL
IS
H
19
4.
After the procedure, the X-NIR
™
analyser shows the result of the optical test:
The internal optical unit of X-NIR
™
is OK!
It is now possible to proceed with execution of the analysis.
Press ENTER or RESET to exit and continue.
5.
If the optical test procedure is unsuccessful, the X-NIR
™
analyser displays the following
message:
ATTENTION!
Problems with the internal optical unit.
Refer to the following troubleshooting procedure.
Press ENTER or RESET to exit and continue.
In any case it is possible to save the optical spectra on the USB key by pressing the function
key
, then send the data to the Support Service dinamica generale for further
inspection.