Digitimer Ltd – D188 Operator’s Manual 

 

Version 1.2 

 

18 

 
 

4:8 Digital Control 

If eight digital outputs are not available, the operator can choose to use four digital outputs, defining a 4-
bit binary string that the D188 receives through pins 1 – 4 (D0 – D3).  As with 1:1 mode, channels remain 
active while the appropriate inputs are held TTL high. 

 

Example Digital Input Configurations: 

Channel 2 Active 

Channel 4 Active 

Channel 5 Active 

Channel 8 Active 

Pin 1 = 0 (TTL low) 

Pin 1 = 0 (TTL low) 

Pin 1 = 1 (TTL high) 

Pin 1 = 0 (TTL low) 

Pin 2 = 1 (TTL high) 

Pin 2 = 0 (TTL low) 

Pin 2 = 0 (TTL low) 

Pin 2 = 0 (TTL low) 

Pin 3 = 0 (TTL low) 

Pin 3 = 1 (TTL high) 

Pin 3 = 1 (TTL high) 

Pin 3 = 0 (TTL low) 

Pin 4 = 0 (TTL low) 

Pin 4 = 0 (TTL low) 

Pin 4 = 0 (TTL low) 

Pin 4 = 1 (TTL high) 

 

Delay/De-bounce 

The Delay/De-bounce setting allows the operator to introduce an adjustable delay in switching which can 
be as long as 1000 milliseconds.  This ensures that the voltage at any of the digital inputs has stabilised 
before the D188 makes changes to the active electrode pair.  This may be of use if mechanical switches 
are used to control the D188, as “switch bounce” could be interpreted by the D188 as multiple rapid 
changes between TTL low and high, rather than a single state change. 

 

Minimum Switching Times & Stimulus Pulse Duration 

It is important to appreciate that the electrode switching process is not instantaneous.  It takes several 
hundred microseconds for the D188 to switch channels in response to a change in the TTL state of the 
digital inputs.  We recommend that stimulus delivery and digital switching protocols are devised which 
allow for a 1ms pause in stimulation following a digital input state change.  If this pause is not adhered to, 
stimulation may occur in multiple channels during the 1ms transition period. 

 

The recommended 1ms pause, following changes in the digital inputs, is indicated by the red blocks in the 
illustration above. 

Содержание D188

Страница 1: ...Digitimer D188 REMOTE ELECTRODE SELECTOR OPERATOR S MANUAL For Research Use Only Digitimer is a registered trade mark of Digitimer Limited...

Страница 2: ...erator s Manual Version 1 2 2 Digitimer Limited 37 Hydeway Welwyn Garden City Hertfordshire AL7 3BE UK Tel 44 0 1707 328347 Fax 44 0 1707 373153 E mail sales digitimer com technical digitimer com Webs...

Страница 3: ...dware 10 Section 3 Software Overview 12 Introduction 12 Standby 12 USB Manual Control Mode 13 1 1 Digital Control Mode 13 4 8 Digital Control Mode 13 Delay De bounce Setting 13 LED Control 13 Persiste...

Страница 4: ...l Control 18 Delay De bounce 18 Minimum Switching Times Stimulus Pulse Duration 18 Digital Control Voltage Requirements 19 Using the API for Third Party Software Control 19 Example Applications 19 App...

Страница 5: ...le digital inputs applied to the rear panel socket on the D188 or manually via Windows compatible software supplied A pair of stimulus inputs is provided for connection to a compatible electrical stim...

Страница 6: ...l supplied on USB Stick Digital Communication Cable 1m long 15 way D connector to 10 tinned wires Optional Accessories D185 OC1 4mm Plugs pair of 4mm touch proof plugs compatible with the outputs on D...

Страница 7: ...ckets Rear panel Connections USB Input Type B Socket for Win PC software control setup Power Supply Input Socket Digital Communication Socket 15 way Female D Connector Indicators 8 Green LEDs One per...

Страница 8: ...Digitimer Ltd D188 Operator s Manual Version 1 2 8...

Страница 9: ...rsion 1 2 9 Section 2 D188 Installation Hardware Overview D188 Front Panel D188 Rear Panel LED Indicator Electrode Connection Sockets 8 pairs Stimulator Input Sockets one pair USB B Socket 15 way Digi...

Страница 10: ...ck Note that pins 11 to 15 are unused A 5V pin is made available for anyone wishing to use this for button control of the D188 and is not needed for digital control Installing the D188 Control Softwar...

Страница 11: ...allation is complete double click the D188App icon on the desktop to run the control software 5 If the software detects one or more D188 s the device ID s serial numbers will be shown in the top left...

Страница 12: ...ch D188 is being managed by the number keys at a particular time The D188 has three modes of operation and a standby setting Modes are selected via a left mouse click on the mode button of choice A gr...

Страница 13: ...tal outputs are not available the operator can choose to use four digital outputs defining a 4 bit binary string that the D188 receives through pins 1 4 D0 D3 See 4 8 Digital Control for more detail D...

Страница 14: ...Digitimer Ltd D188 Operator s Manual Version 1 2 14...

Страница 15: ...p is connection of the D188 to the source of stimulation the stimulation electrodes and the digital trigger source A schematic of a typical setup is given below The diagram above illustrates how a sin...

Страница 16: ...end and 1 5mm touch proof female connectors at the D188 end Connecting to a Digitimer DS2A DS3 or DS4 Stimulator As our non human research stimulators are fitted with 2mm touch proof sockets so we off...

Страница 17: ...the operator to test and optimize the physical connections between stimulus source computer the D188 and the subject whilst also ensuring that the stimulator settings and electrodes are appropriate f...

Страница 18: ...ng as 1000 milliseconds This ensures that the voltage at any of the digital inputs has stabilised before the D188 makes changes to the active electrode pair This may be of use if mechanical switches a...

Страница 19: ...velopers we provide a sample C example program within a zip file D188 VS C Example zip located in C Users username Documents Digitimer Limited Example Applications Below we include an example applicat...

Страница 20: ...ited 37 Hydeway Welwyn Garden City Hertfordshire AL7 3BE UK Tel 44 0 1707 328347 Fax 44 0 1707 373153 E mail sales digitimer com technical digitimer com Website www digitimer com File Reference N Docs...

Отзывы: