1 Product introduction
ThermoVu DTM-700 infrared thermal imaging temperature measurement and screening system is
composed of human body temperature camera, human body temperature measurement blackbody
and temperature measurement and screening software. Its main feature is non-contact temperature
measurement, which is suitable for large-scale screening in the scene of train station, airport, bus
station, hospital and school with large flow of people, so as to realize temperature anomaly warning.
It can be widely used for medical temperature measurement, real-time temperature information
transmission, accurate detection, rapid diagnosis, etc.
The core purpose of the temperature measurement and screening system is to screen the target of
fever personnel from the normal temperature population, and then check it again through the
medical temperature measurement instrument.
2 Site Survey and Commissioning Guide
ThermoVu DTM-700 infrared thermal temperature measurement and screening system belongs to
precision temperature measurement and imaging equipment, it has the following requirements for
engineering exploration application.
2.1 Scenario requirements
⚫
The ambient temperature should be kept relatively constant to avoid obvious changes in the
body surface temperature of the tested personnel;
When the ambient temperature is within 16-32
℃
, the equipment supports accurate
temperature measurement, and the temperature deviation value is
±
0.3
℃
;
When the ambient temperature exceeds 16-32
℃
, the equipment will not be able to
guarantee the accuracy of temperature measurement;
Indoor scene: It is recommended to use diversion belt (L-type or Z-type)
Outdoor scene: outdoor scene is not recommended, if necessary, it must be used with tent /
shelter / sealed shed + diversion belt (L or Z type)
According to the actual application scenario, it may be necessary to adjust the ambient
temperature by adding air conditioner or electric heater;
⚫
The installation position of the camera shall be well illuminated to avoid backlighting;
⚫
The camera monitoring area shall not be interfered by strong heat sources such as strong light
or sunlight;