Attention
:
1.
The calibration point must be calibrated on the place of 5m position, with the reference
person or obvious reference as the calibration point.
2.
The calibration point must be close to the diagonal of the screen.
3.
If the equipment position does not move after one calibration, it is not necessary to
calibrate again.
3.5.3 Blackbody configuration
➢
Enter “System para”-> “Blackbody configuration” interface
➢
First, select the location of the blackbody, and include the whole round target of the blackbody
in the blackbody box.
➢
Then right click the blackbody box to turn on the blackbody, and the blackbody temperature
should be 37
℃
consistent with the actual blackbody box display temperature, and the specific
emissivity should be set to 1.
➢
Click "Apply" to complete the setting.
Attention
:
1.
After the blackbody is powered on, check if there is a temperature display on the
blackbody rear panel. The default setting is 37
℃
, it means that the blackbody is working
normally when the upper and lower temperatures are the same (or no more than
±
0.2
℃
),
2.
The round target of the blackbody should be kept facing the camera.