Your Cardio Care unit is warranted to be free from defects in materials 
and workmanship appearing within 1 year from the date of purchase, 
when used in accordance with the instructions provided. The above 
warranties extend only to the original retails purchaser.

According manufacture date, please see product’s CE label SN number,
the SN coding principles are as below.

SN Coding principles

The first to fourth codes 0009 are the product part number, A9800009, 
last four codes.
The fifth to sixth codes “YY” are the last 2 codes of A.D.
The seventh code “M” is month (1~9, Oct=A, Nov=B, Dec=C) 
The eighth to ninth code “DD” are manufacture date 
The tenth to fourteenth code “XXXXX” are serial number 
The fifteenth code “R” is the version of the product (A9800009)

Sings and Symbols

Problem

Cause

Solution

Nothing happens when 
you press the power 
button

Measurement finishes 
but result is abnormal.

The batteries are worn 
out.  The batteries are 
inserted incorrectly. 

The electrodes are not 
making good contact 
with your finger or 
chest. 

Your body is too tense. 

Electrode contact was 
broken before 30 
seconds of data has 
been recorded. 

The chest electrode has 
been applied over 
clothing. 

Your skin is dry. 

The ECG signal is too 
weak. 

Replace both batteries 
with new ones. Insert the 
batteries with their 
polarities correctly 
aligned.

Check the instructions 
and repeat the 
measurement.

Relax and repeat the 
measurement.

Make sure that your 
finger and chest maintain 
contact with the 
electrodes for the full 30 
seconds of the 
measurement.

Apply the chest electrode 
to bare skin.

Moisten your skin with a 
damp towel or wash your 
hands in advance, but 
don’t use the unit in wet 
hands.

Make sure that the 
electrodes are applied 
correctly. However, for 
some individuals 
measurement may not be 
possible.

18

19

Warranty Information

Manufacture Date

Troubleshooting

Type B Equipment

Caution/Warning, Consult accompanying 
documents

CE-European Conformity. This device 
conforms to the standards for products 
sold in the European Economic Area.

Interference may occur in the vicinity of 
equipment marked.

An item that poses no known hazards in 
all MR environments.

Consult instruction for use

Product Name: Cardio Care

Model No: ECG-10X

Power: Battery AAAx2  3Vdc

Manufacturer: digiO2 International Co., Ltd.

4F-13, No. 79, Sec. 1, Hsin Tai Wu Rd., Hsi-Chih, Taipei Hsien 221, Taiwan

S/N:

0009YYMDDXXXXXR

0120

FCC ID : X58-ECG10X

Содержание ECG-10X

Страница 1: ...ECG EKG Recorder ...

Страница 2: ...g Instructions 04 Power on off Measurement Historical Record 07 Review Measurement Record Delete Record Setting 09 Date and Time Alarm Bluetooth 10 Bluetooth Upload record to PC server 11 USB Upload record to PC server 13 Important Safety Note 15 Maintenance and Storage 17 Troubleshooting 18 Warranty Information 19 Manufature Date 19 Specifications 20 ...

Страница 3: ...nd period These ECG readings can monitor your condition more easily Cardio care unit is designed for prescription purpose record store and display the ECG waveform and heart rate that it can help doctor to know the patient s condition This unit is intended for use on the adult population over 20 years old NOTE Please note that while the measurement from an ECG EKG recorder is an useful guide for y...

Страница 4: ... with a beep sound after 5 seconds Place the unit with the back end electrode against the spot slightly below your chest The measurement will stop after 30 seconds of recording automatically and is accompanied with a beep sound The heart rate will be shown on the page and press D The unit will ask automatically if you want to upload the record to server press to upload or to skip E After selection...

Страница 5: ...t Movement including talking coughing or sneezing during measurement can affect the measurement results If the position of your finger or arm is inappropriate during measurement the results may be inaccurate so be sure to check their positions before taking a measurement After taking measurement disinfect the electrodes by wiping them with a soft cloth or cotton swab moistened with disinfectant al...

Страница 6: ...rd Setting Setting page can be accessed directly through Menu selection Date and Time Press to set date and time and to confirm Press to return to the previous step Press to decrease the number Press to increase the number Alarm Press to set the alarm time Press to decrease the number Press to increase the number Please press to select ON OFF and to confirm Press to return to the previous step 08 ...

Страница 7: ...o2 com before uploading record to PC server Go to http www digio2 com Press Download Press Healthcare Device Press My Cardio Care Download and install My Cardio Care to your PC 2 Setting Device Name 3 Connect Bluetooth When you see the Bluetooth device has been connected then key in 0000 as passkey 4 1 Go to the device s Historical record page Bluetooth Upload record to PC server Users are allowed...

Страница 8: ...KG waveform 1 Download My Cardio Care Please download and install My Cardio Care from our website http www digio2 com before uploading record to PC server Go to http www digiO2 com Press Download Press Healthcare Device Press My Cardio Care Download and install My Cardio Care to your PC 2 Turn Off Bluetooth Make sure the Bluetooth of Cardio Care is turned off Press to select ON OFF and to confirm ...

Страница 9: ...n their eye and who have an aneurysm clip in their brain Do not attempt self diagnosis or self treatment based on the measurement results and analysis Always consult your doctor Self diagnosis or self treatment may lead to deterioration in your condition Do not use this unit on people with sensitive skin or allergies Do not attempt to use the unit over or through clothing Do not attempt to use the...

Страница 10: ...ystem does not provide sufficient insulation to mains and earth however the equipment can be safely connected to other IEC EN 60601 1 approved equipment that has all its connectors intended for connection to this unit separated from earth by basic insulation Do not use this product in combination with non IEC EN 60601 1 approved computers adaptors Any ECG EKG record erased from the unit cannot be ...

Страница 11: ...onds of data has been recorded The chest electrode has been applied over clothing Your skin is dry The ECG signal is too weak Replace both batteries with new ones Insert the batteries with their polarities correctly aligned Check the instructions and repeat the measurement Relax and repeat the measurement Make sure that your finger and chest maintain contact with the electrodes for the full 30 sec...

Страница 12: ...ence Be sure to read the Instruction Manual for correct installation and use Also be sure to read the Instruction Manuals for all other electronic equipment nearby ECG 10X 1 5V AAA Battery 2 10mm mV 5 10M 2mV 0 005 40 Hz 95 dB 12 bits 256 samples sec 30 sec 125x64x27 144g 20 C 50 C 10 C 40 C 25 95 USB Bluetooth USB Bluetooth 60mm x 32 5mm LCM display panel ECG EKG Specifications Power Supply Stand...

Страница 13: ... by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help This device complies with part 15 of the FCC Rules Operation is subject to the follow...

Отзывы: