SM500MK4 Service Manual
43
4) In the Hyper Terminal software to select [
File
]
Æ
[
Properties
] to open
the properties setting.
5) Go to Settings Option use the
button to change the setting follow
by below highlight as shown.
Emulator:
ANSIW
Telnet terminal ID:
VT100
Click the [
OK
] button to exit and go to downloading page.
6) Switch on the scale power, the screen will shown [
Press any key to
enter SM300/500 Basic Boot-loader
], then press the keyboard
[
Enter
] key button 2 times.
7)
a) On the SM300/500 Basic Boot-loader Main Menu will show
[Enter erase start address as a 32-bit hex value e.g. AABBCC].
b) Type in the number [
100800000
].
Note: When type in the first number “
1
”, it will automatically convert to “
H’
“ symbol.
Remarks:
I.
After next step on the screen shown [Enter erase end
address as a 32-bit hex value e.g. AABBCC]. Around 3
seconds lead time to continue, if not it will shown the
[Timed out waiting for address….].
[Invalid end address – aborting].
II.
You can type in the number [
100800000
] again and to
continue.
All manuals and user guides at all-guides.com
all-guides.com
Содержание SM-500MK4
Страница 9: ...SM500MK4 Service Manual 6 2 1 3 Bench 462 154 All manuals and user guides at all guides com ...
Страница 58: ...55 10 APPENDIX 10 1 Block Diagram 10 1 1 Bench Type All manuals and user guides at all guides com ...
Страница 59: ...56 10 1 2 Touch Screen Pole Type All manuals and user guides at all guides com ...
Страница 60: ...57 10 1 3 Elevated Type All manuals and user guides at all guides com ...