3.7. Set-Point Setting (SPEC8 BIT0 = 1)
Indication Flag: B - Bright S - Sparkling
1-ZERO 2-NET 3-STD 4-U.C 5-SAF 6-SAC 7-S1 8-S2 9-S3 10-S4
Indicator
Operation
Key
Weight
1 2 3 4 5 6 7 8 9 10
Remarks
Weight Stable
0.0000
kg B
B
Weighing mode
Enter into
setting mode
[Mode]
[Mode]
SET-1
Set for Point 1 *1)
Confirm [Enter]
0.000
0
kg
B
Set value
Value change
[Shift]
[INC]
0.00
5
0
kg
B
[Shift] key to select
the digit, [INC] key to
change the value of
digit selected *1)
Confirm
[Enter] SET-2
Set-Point
1
saved
Point 2 setting
[Enter]
0.000
0
kg
B
Set value *2)
Value change
[Shift]
[INC]
0.0
1
00
kg
B
Same method as
point 1 setting *2)
Confirm
[Enter]
SET-3
kg
Point 2 saved *3)
Confirm [Enter]
0.000
0
kg
B
Set value *3)
Value Change
[Shift]
[INC]
0.0
1
00
kg
B
Same procedure as
point 1 & 2 setting *3)
Confirm
[Enter]
SET-4
kg
Point 3 saved *4)
Confirm [Enter]
0.000
0
kg
B Set value *4)
Value change
[Shift]
[INC]
0.0
1
00
kg
B
Same method as
above *4)
Confirm
[Enter] 0.0000 kg B B
Point 4 saved and
back to weighing
mode *4)
Note 1: The indicators S1, S2, S3, S4 will be lighted when setting the corresponding Set-Point.
Note 2: When the weight exceeds the value of Set-Point 1, the indicator S1 will light.
When the weight exceeds the value of Set-Point 2, the indicator S1 and S2 are light.
When the weight exceeds the value of Set-Point 3, S1, S2 and S3 are light.
When the weight exceeds Point 4, S1, S2, S3 and S4 are light.
03 Main Operation