DI-5000_DP-5000_DPS-5000 Service Manual Edition 2
45
5) Close back the printer and touch
[FEED]
button, and check
the
[Under Printing]
and
[Printer Busy]
check box will
appear a tick mark short a while.
6) When the
[Thermistor Warning ON]
check box appear the
tick mark, that mean the Thermal Head over heating 80
°
C,
once the Thermal Head cool down lower then 60
°
C, the tick
mark will disappear.
7) When the
[Thermal Head Problem]
check box appear the
tick mark, that mean the thermal head have detected the
printing diagnostic check have problem like dot missing.
Note:
This function required the Printer Spec
[143A026]
Thermal Head
Diagnosis is set to
[Feed]
or
[Feed + Dynamic]
.
7.1.2 Sensor Calibration
Procedure
Picture
1) In Registration mode, select
[MENU]
[MAINTENANCE]
[MAINTENANCE]
[PRINTER]
[PRINTER TEST]
to go to
printer test mode.