![Digi-Sense 69200-00 Скачать руководство пользователя страница 61](http://html1.mh-extra.com/html/digi-sense/69200-00/69200-00_instruction-manual_2496809061.webp)
60
Specifications
Operating Temperature
(for stated accuracy) ........................ 18
°
C to 28
°
C (64
°
F to 82
°
F)
Operating Temperature
(useful range) ................................... 0
°
C to 40
°
C (32
°
F to 104
°
F)
Storage temperature ........................ –40
°
C to 60
°
C (–40
°
F to 110
°
F)
Humidity (non-condensing) .................... 10 to 90%
Altitude ................................................. Less than 2000 m
Accuracy
Thermocouple types E,J,K,N,T
>–150
°
C ...................................
±
(0.1% of reading
±
0.5
°
C)
±
(0.1% of reading
±
0.8
°
F)
<–150
°
C ...................................
±
(0.25% of reading
±
1
°
C)
±
(0.25% of reading
±
2
°
F)
Thermocouple types B,R,S ..............
±
(0.1% of reading
±
2
°
C)
±
(0.1% of reading
±
4
°
F)
Resolution .............................................. 0.1
°
or 1
°
(1
°
above 999.9
°
)
(1
°
below –150
°
C, –238
°
F)
Scales
.................................................
°
C /
°
F / K /
°
R
Isolation
thermocouple to thermocouple ........ 250 VRMS
thermocouple to rear panel .............. 250 VRMS
Thermocouple connections .................... Two-blade female mini-ANSI
Display update rate ................................ 3 seconds
Thermocouple read time ........................ 4 seconds to 60 minutes
Print time ................................................ 4 seconds to 60 minutes
Battery
................................................. Lithium BR1225, RAYOVAC
®
or
PANASONIC
®
only
Display ................................................. LCD
12 0.7 in variable characters
30 fixed word/icons
Alarm level ............................................. 50 dBa
Power source
Line voltage ..................................... 10–28V DC (non-polarized)
@ ~ 300 mA
9–20V AC 50–400 Hz
@ ~ 500 mA