Front panel displays and controls continued
How to set trigger type
How to set trigger type
A trigger (contact between pins 1 and 2 of the trigger) can be set to:
•
pulse print (print one record when trigger contact is closed).
•
S/S (start/stop) PRINT (equivalent to toggling PRINT on while
trigger contact is closed).
•
pulse STORE (equivalent to pushing STORE to log one record).
•
S/S (start/stop) STORE (equivalent to pushing LOG to continu-
ously log records while trigger contact is closed).
+ -
A L A R M
E X I T
MENU
1
S T O R E
2
L O G
3
R E C A L L
4
P R I N T
5
M A X
6
M I N
7
A V G
8
H O L D
9
S C A N
1b
1, 3b
3a
1a
2
0
CLEAR
The default setting is OFF.
To set or adjust the trigger type:
1. Push MENU until TRIGGER 1a is displayed and the current
setting 1b (OFF, STORE or PRINT) flashes.
2. Push the Up or Down arrow keys to scroll between OFF, STORE
and PRINT.
3a. Push EXIT to save the setting and return to scanning, or
3b. Push MENU to switch to the next setup topic (set trigger mode).
BA 32-36