Instructions Manual
34/50
Rev 2.0 of 02/2016
4.3
TEST TUBE TYPES
VES-MATIC CUBE 30 is capable of analysing various types of test tube (Figure 4.2.1) in the same
cycle, however, there are limitations for rubber-capped test tubes that require the plate reducer in order
to be processed.
If rubber-capped test tubes are used, the
reducer disc must be fitted on the plate.
The disc must only be used for rubber-
capped test tubes (see paragraph 4.3).
Fig 4.3.1
Fitting the plate reducer
The reducer has been designed specifically for laboratories that need to process different types of test
tubes, among which rubber-capped, so that it can be very easily fitted/removed several times a day.
Step 1
Place the reducer on the plate in such a way that the 3
locking pins slip into the dedicated housings (Figure 4.3.2)
and that the groove in the pin head is horizontal with respect
to the plate numbering.
Step 2
Push the pin down and turn it 90 degrees.
Step 3
Repeat the operation for the other two pins.
Remember that the reducer disc is to be fitted only for use with rubber-capped test tubes
and must be removed for use with Vacutainer® (BD), Vacuette® and others test tubes.
Fig 4.3.2
Fig 4.3.2.a
Fig 4.3.2.b