C U B E 3 0 T O U C H - U S E R M A N U A L
50/56
Rev 1.1 07/2018
Anomalous test
tube
Err (15)
This may be due to
significant sedimentation
with a low volume of blood.
Check for the presence of
clumps or clots. Try to repeat
the test.
Exam not available
Reloading required
Zero tests remaining.
Follow the instructions for
reloading the instrument with
a new test device, as
described in chapter 4.
USB device not
found
The instrument does not
recognize the USB device.
Remove and reinsert the
device or replace it.
Caution! “N” tests
remain
The number of remaining
tests is less than 50.
Be sure to reload the device
as soon as possible with a
new test device
Used TEST DEVICE
The test Device in use is
empty.
Obtain a new test device.
.