Troubleshooting 1 5
...fioNAeneermove,.........teranorso.
Troubleshooting
The add plug-in command only works if there is one new, cleared, or deleted plug-in
inserted into the GP700. If there are two or more new plug-ins inserted when you begin the
add procedure, the attempt
will
usually fail and give an error message. It is possible the add
procedure will appear to work, but will not complete successfully. In this case, when you
exit the add procedure and reset the instrument, the screen will not show a display box for
the new devices.
Although the plug-ins were not added successfully, the GP700M may have created errors
in the configuration records in the GP700M as well as in each of the plug-ins. To correct a
multiple add error, execute the clear plug-in command
(CLRP)
and re-install the plug-ins
one at a time.
Figure 11: Add
Plug-in Error Screen
Operation failed!
ENTER to try again.
ESC to stop.
HELP twice to exit.
Attempting to add a plug-in without inserting the physical cartridge will also result in an
"operation failed" error. Insert a plug-in and press
ENTER
to try again.
91507 Revision A3
March 11, 2005