TROUBLESHOOTING / 99
TROUBLESHOOTING
If the system locks, press and hold the
On/Off
button on the front of
machine until system turns off or unplug instrument power. Wait a few
seconds and then power on.
The GAC
®
2500 contains a microprocessor to control grain sample
measurements and self-checks that determine the integrity of the internal
electronics. If any limit is exceeded or malfunctions occur an error code
displays with a recommended resolution.
Figure 94
Error Screen Example
ERROR MESSAGES
Error messages display when an abnormal event occurs. An error can be
acknowledged by pressing the
Initiate
button. The table provides an
overview of error codes with potential causes and corrective actions. For
any failure that persists, contact DICKEY-john Europe at +33 141 19 21 80
for service or DICKEY-john US Technical Support at 1-217-438-3371.
Error Detected
Error Detected
Error: 5
INVALID GRAIN CALIBRATION FILE
An invalid calibration file was
selected.
Re-install the selected calibration
file.
If error persists contact technical
support.
c:\gac2500\product\Durum-Bad.cal
Содержание GAC 2500
Страница 4: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 IV ...
Страница 20: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 16 SECURITY ...
Страница 24: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 20 INITIAL SETUP ...
Страница 28: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 24 NAVIGATION ...
Страница 66: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 62 SETUP ...
Страница 88: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 84 GRAIN CALIBRATIONS ...
Страница 90: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 86 DIAGNOSTICS ...
Страница 102: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 98 MAINTENANCE ...
Страница 104: ...OPERATOR S MANUAL GAC 2500 Model INTL 11001 1655A 201603 100 TROUBLESHOOTING ...