TH800/02/03/05

Temperature & Humidity Recorder

www.DicksonData.com • 1-800-323-2448 • Fax 1-800-676-0498

2. 

   Both pens will move to the outer edge of the chart, 

and  then  the  shorter  red  pen  will  move  to  the 

current  position.  Use  the 

Adjust-Up 

and 

Adjust-

Down

  buttons  to  move  the  pen  to  the  maximum 

temperature  line  on  the  chart.  Once  the  shorter 

red pen is set, the 

Pen Home

 button will move the 

shorter red pen to the outer edge of the chart and the 

longer blue pen will move to its current position. The 

Adjust-Up

  and 

Adjust-Down

  buttons  will  adjust 

the longer blue pen position. Once the pen is set, the 

Pen Home

 button will move the longer blue pen to 

the outer edge of the chart and the shorter red pen 

will move to its new position. Each time pen home is 

pressed the two pens will switch positions (moving 

the  most  recently  adjusted  position)  until  the 

On/

Off

 button is pressed to exit pen adjust mode.

Note:

  Pressing the 

On/Off

 button will exit the adjustment 

and  return  the  unit  to  normal  operation.  Each  time  the 

Pen Home

 or 

On/Off

 button is pressed, the adjustment 

currently in progress will be stored. The unit will take one 

minute to exit 

Pen Adjust Mode

 once the 

On/Off

 button 

is pressed.

TROUBLESHOOTING

Why isn’t the chart keeping time or running slow? 

•  The  chart  may  be  “hung  up”  or  restricted,  possibly 

caused by a rip on the outer edge of the chart or the 

chart hub, or the chart may be caught between the 

arm and pen arm platform. 

•  Incorrect chart installed for the selected chart speed. 

Why did the chart stop turning?

 

• Chart hung up or restricted, (ripped chart) 
•  Unit may be “locked up”? This can be confirmed by 

pressing  any  of  the  buttons  on  the  keypad.  If  the 

unit is locked up there will be no response to button 

presses  and  the  Chart  Recorder  may  appear  to  be 

working,  but  the  readings  won’t  change.  Also,  the 

chart will not rotate. Remove power and battery, and 

then re-power. 

Why don’t the display and chart match?

 

•  Dip switches are set for a specific range, but using 

chart for another range, or vice versa. 

• Pen(s) not inserted on pen arm(s) all the way 
•  To adjust pen to match chart see 

Pen Adjust

 in the 

Need to Know

 section of this manual. 

Note:  TH800 

does not have a display.

Why does calibration seem to be off?

 

•  What is the tolerance of the unit it’s being compared to? 
•  It’s  ok  if  the  unit  is  within  the  sum  of  the  two 

tolerances. 

•  Has  an  outside  calibration  house  attempted 

calibration? It may not have been adjusted properly. 

•  Walk  through  calibration  adjustment,  found  in  

the manual

Why won’t the battery backup work?

 

• Are there good batteries in the Chart Recorder?
•  Keep in mind that battery backup varies significantly 

depending on temperature, pen movement and chart 

rotation speed. 

•  If there was a momentary power outage (brown out), 

the unit may not have had enough time to recognize 

this and switch to 

Battery Mode

. (The Chart Recorder 

may lock up or shut off in this situation) This situation 

may  also  occur  if  the  unit  is  plugged  into  an  outlet 

that shares the circuit with other machinery that has 

phase motors or compressors that cycle periodically. 

As these other machines cycle, they momentarily have 

a high current draw, therefore drawing from the unit. 

Note:  TH800

 does not have a battery backup.

Why won’t the chart recorder respond to changes?

 

•  Is  the  unit  “locked  up”?  This  can  be  confirmed  by 

pressing  any  of  the  buttons  on  the  keypad.  If  the 

unit is locked up there will be no response to button 

presses  and  the  unit  may  appear  to  be  working, 

but  the  readings  won’t  change.  Also,  the  chart  will 

not  rotate.  Remove  power  and  battery,  and  then 

re-power. 

Why won’t the unit power up? 

•  Remove  the  batteries  and  power  adapter  for  a 

minute or two, this will reset the unit. The unit should 

respond when the adapter is plugged in.

Warranty

Dickson warrants that this line of instruments will be 

free from defects in material and workmanship under 

normal use and service for a period of twelve months 

after delivery.
This  warranty  does  not  cover  routine  calibration  and 

battery replacement.
For 

Specifications

  and 

Technical  Support

  go  to 

www.DicksonData.com