TH800/02/03/05

Temperature & Humidity Recorder

www.DicksonData.com • 1-800-323-2448 • Fax 1-800-676-0498

Getting Started

Default Recorder Settings

  7-day chart rotation

  +32 to 120F (TH800/02)

  -20 to +120F (TH803/05)

Quick Start

1. 

 Remove the protective pen caps. 

2. 

  Insert batteries and plug in AC adapter.  

(See Figure 3.) (

TH8

 requires 4 AA batteries only.) 

The unit will power on.

3. 

  The instrument will move pen(s) to the appropriate 

readings.

4. 

  Install the chart that matches the Dip Switch settings

 Press the 

Pen Home

 key to move the pens to the 

outside of the chart. The pen is automatically raised 

off the chart.
Remove the old chart, place the new chart on the 

Chart Hub

 being certain that the edge of the chart 

slides  under  the 

Chart  Guide  Clips

  located  at  the 

outside of the chart.

5. 

  

Set the appropriate time.

 There are two ways to 

adjust the chart and set the appropriate time:

a.

  Set the chart time manually by inserting a coin into 

the groove in the chart hub and turning clockwise 

until the correct hour (and day if applicable) on the 

chart is referenced to the timing arrow (just to the 

right of the pen tip on the dial). 

This feature should be used for fine adjustments only. 

b.

  To adjust the chart time, press and hold 

Adjust-Up

 and 

Adjust-Down

 buttons located 

on the back of the unit next to the dip switch (see 

Figure 3). 

The  green  LED  will  blink  rapidly  for  about  five 

seconds,  then  the  LED  will  remain  solid  green. 

While in this state the 

Adjust-Up

 button will move 

the  chart  backward  (counter  clockwise)  and  the 

Adjust-Down

 button will move the chart forward 

(clockwise). 
Rotate  the  chart  until  the  correct  hour  (and  day  if 

applicable) on the chart is referenced to the timing 

arrow. Once you have set your chart, press the 

Pen 

Home

  button  to  exit 

Chart  Adjust  Mode

.  The  unit 

will take one minute to exit the 

Chart Adjust Mode 

once 

Pen Home

 is pressed.

 Press 

Pen  Home

  to  move  the  pens  back  onto 

the chart.

6. 

  Place  the 

TH8

  recorder  on  a  flat  vibration-free 

surface. Be sure it is in a vertical position and level. 

For  best  performance  and  longevity,  the  location 

should  be  a  clean  environment,  free  from  dust 

and  corrosive  fumes.  Do  not  exceed  temperature 

specifications.

Wall Mount

  Keyhole slots are provided on the 

TH8

 

recorder for wall mounting.

Display Symbols  TH8

Figure 1

MAX

MIN

F or C

RH

+

-

Low battery 

indicator

Current reading

or

Flashing “ALRN” when

in alarm condition

For setting alarms only

Battery enabled