background image

KT800/802/803/804/809/856  

Remote Sensing  

Temperature Recorder

www.DicksonData.com • 1-800-323-2448 • Fax 1-800-676-0498

pens  on  a  two-pen  unit  and  stores  the  current  

adjustment value. 

3. 

   When  calibration  is  complete,  press  the 

On/Off 

button. Calibration is stored in memory even after 

you  turn  the  unit  off.  User  calibration  information 

will not be lost if AC power fails.

Note: 

After two hours, if no buttons are pressed, the 

unit will time out of user calibration and resume normal 

operation.  The  user  calibration  is  stored  separately 

from the factory calibration. If you wish to cancel out 

your user calibration, simply enter 

Calibration Mode

 and 

toggle  through  the  steps  without  adjusting  displayed 

readings. Exit by pressing the 

On/Off

 button. You have 

now restored factory calibration.

Note:

  It  is  recommended  that  you  use  a  controlled 

chamber  when  calibrating  temperature  and  humidity. 

Calibrating  in  an  open  room  is  not  recommended  as 

humidity can vary greatly within a very small area.

NEED TO KNOW

KT809 Water Resistant Enclosure

The

 KT809 

features an IP66 enclosure. In order to maintain 

the IP66 rating the battery/switch cover, on the back of the 

unit, must be screwed down tightly at all points. The door 

must be closed tightly using the two side latches. The K-TC 

probe must be connected using the external connector and 

the connector/jack connection must be completely sealed 

using waterproof tape.
It is strongly recommended that the case door gasket be 

greased  occasionally  using  a  standard  silicone  rubber 

grease. The more the door is opened and closed, the more 

frequently grease should be applied.
The Door and Battery Cover gaskets should maintain the 

IP66 rating for 1 year under normal use. To replace gaskets, 

contact 

Dickson Customer Service.

 

KT800/09 Pen Movement

The pen movement for the 

KT800/09

 is slower than on the 

AC powered models in order to conserve battery power. 

In 24-hour mode, the pen will move every 1 minute and 30 

seconds. In 7-day mode, the pen will move every 8 minutes 

and 30 seconds. 
The recorder takes temperature readings every 50 seconds. 

An average of the readings taken between pen movements 

is what is drawn on the chart. 

Pen adjust

Pen adjust may be needed if the pen(s) and display do not 

match.

1. 

   While the unit is on, press and hold both the 

Pen 

Home

 button and the 

On/Off

 button (making sure to 

press the 

Pen Home

 button first so the unit will not 

turn off). The led will blink rapidly between amber 

and green for about five seconds. The LED will then 

turn  green.  Release  the 

Pen  Home

  and 

On/Off 

buttons. The LED will flash amber and green for one 

second and the LED will turn off.

2. 

 

 One-pen models 

The red pen will move to the outer 

edge of the chart and is ready to be adjusted. Use 

the 

Adjust-Up

 and 

Adjust-Down

 buttons to move 

the  pen  to  the  maximum  temperature  line  on  the 

chart.  Once the pen is set, the 

Pen Home

 button 

will move the pen to reaffirm the new pen position. 

Press the 

On/Off

 button to exit 

Pen Home Adjust

.

3. 

   

Two-pen models

  Both pens will move to the outer 

edge of the chart, and then the shorter red pen will 

move  to  the  current  position.  Use  the 

Adjust-Up 

and 

Adjust-Down

  buttons  to  move  the  pen  to 

the  maximum  temperature  line  on  the  chart.  Once 

the  shorter  red  pen  is  set,  the 

Pen  Home

  button 

will move the shorter red pen to the outer edge of 

the chart and the longer blue pen will move to its 

current position. The 

Adjust-Up

 and 

Adjust-Down

 

buttons  will  adjust  the  longer  blue  pen  position. 

Once the pen is set, the 

Pen Home

 button will move 

the longer blue pen to the outer edge of the chart and 

the  shorter  red  pen  will  move  to its new  position. 

Each  time  pen  home  is  pressed  the  two  pens  will 

switch positions (moving the most recently adjusted 

position) until the 

On/Off

 button is pressed to exit 

pen adjust mode.

Note:

  Pressing the 

On/Off

 button will exit the adjustment 

and  return  the  unit  to  normal  operation.  Each  time  the 

Pen Home

 or 

On/Off

 button is pressed, the adjustment 

currently in progress will be stored. The unit will take one 

minute to exit 

Pen Adjust Mode

 once the 

On/Off

 button 

is pressed.

TROUBLESHOOTING

Why isn’t the chart keeping time or running slow? 

•  The  chart  may  be  “hung  up”  or  restricted,  possibly 

caused by a rip on the outer edge of the chart or the 

chart hub, or the chart may be caught between the 

arm and pen arm platform. 

•  Incorrect chart installed for the selected chart speed. 

Why did the chart stop turning?

 

• Chart hung up or restricted, (ripped chart) 
•  Unit  may  be  “locked  up”?  This  can  be  confirmed  by 

Отзывы: