KT800/802/803/804/809/856
Remote Sensing
Temperature Recorder
www.DicksonData.com • 1-800-323-2448 • Fax 1-800-676-0498
Pen 1 alarm minimum
Pen 1 alarm maximum
If two pen unit, the additional options are:
Pen 2 alarm minimum
Pen 2 alarm maximum
3.
In order to set pen alarm minimums and maximums,
pressing
Adjust-Up
will increase the alarm value,
and pressing
Adjust-Down
will decrease the alarm
value. There is acceleration if the
Adjust-Up
button
is held down.
Repetitively pressing the
Pen Home
button will
scroll through the five options until the alarm button
is pressed to exit from alarm adjust. Each press of
the
Pen Home
or
Alarm
button will store the new
settings. The unit will take one minute to exit
Alarm
Set Mode
once the
Alarm
button is pressed.
4.
If the alarm is triggered, the LED will show as solid
red and the alarm will sound. Press the
Alarm
button
to silence the audible alarm.
Alarm Relays
The SPST 24V 500mA relay contacts are normally open and
will close on alarm conditions when the alarms are enabled.
Relay contacts are always functional when the alarm is
enabled. The relay will close only during minimum and
maximum alarm conditions. Two pen models have 2 sets
of relay contacts which operate independently. Relays will
not operate under battery only power conditions.
Dip Switch Setup
To set up the
KT8
recorder for your specific application, you
might need to change some of the “Dip Switches”.
The Dip
Switches are located on the back of the unit. (Figure 3)
Use a pen or small screw driver to flip the switches.
Remember to install the correct chart to match the
corresponding Dip Switch setting.
1
2
3
4
5
6
7
8
Slide toggle up for “on” position
Slide toggle up for “on” position
Dip Switch 1
7 & 8 are
not active on
this recorder.
Slide toggle down for “off” position
KT8 Dip Switch Setting Up=On Down=Off
Switches 1–6 For Model
1
2
3
4
5
6
7-Day
All Models
D
D
24-Hour
All Models
D
U
31-Day
All Models
U
U
Fahrenheit
All Models
D
Celsius
All Models
U
-50 to 50° F/C All Models
D
D
U
0° to 100°F/C All Models
D
D
D
0° to 250°F/C All Models
D
U
U
0° to 500°F/C KT802/03/
05/55/56
U
D
D
0 to 150° F/C KT802/03/
05/55/56
D
U
D
Remember to install the correct chart to match corresponding switch setting
DICKSON
•
Addison, IL
•
www.DicksonData.com
KT8 Dip Switch Setting Up=On Down=Off
Switches 1–6 Model
1
2
3
4
5
6
7-Day
KT804
D
D
24-Hour
KT804
D
U
Fahrenheit
KT804
D
Celsius
KT804
U
0 to 1000° F/C KT804
D
D
D
0 to 2000° F
KT804
D
D
U
-20° to +20°F/C
KT804
D
U
D
+20° to +100°F/C KT804
D
U
U
+50 to 150° F/C KT804
U
D
D
Remember to install the correct chart to match corresponding switch setting
DICKSON
•
Addison, IL
•
www.DicksonData.com
Recording Time
All the
KT8
recorders feature user selectable
recording times.
24-hour all models
#1 down
#2 up
7-day all models
#1 down
#2 down
31-day
KT802/03/56
#1 up
#2 up
°F/C
You can record in °F or °C with the
KT8
by using dip
switch #3.
°F
#3 down
°C
#3 up
Note:
Remember to install the correct chart to match
corresponding switch setting.