KT800/802/803/804/809/856  

Remote Sensing  

Temperature Recorder

www.DicksonData.com • 1-800-323-2448 • Fax 1-800-676-0498

Pen 1 alarm minimum

Pen 1 alarm maximum
If two pen unit, the additional options are:
Pen 2 alarm minimum

Pen 2 alarm maximum

3. 

 

  

In order to set pen alarm minimums and maximums, 

pressing 

Adjust-Up

  will  increase  the  alarm  value, 

and pressing 

Adjust-Down

 will decrease the alarm 

value. There is acceleration if the 

Adjust-Up

 button 

is held down. 

Repetitively  pressing  the 

Pen  Home

  button  will 

scroll through the five options until the alarm button 

is pressed to exit from alarm adjust. Each press of 

the 

Pen Home

 or 

Alarm

 button will store the new 

settings. The unit will take one minute to exit 

Alarm 

Set Mode

 once the 

Alarm

 button is pressed. 

4. 

  If the alarm is triggered, the LED will show as solid 

red and the alarm will sound. Press the 

Alarm

 button 

to silence the audible alarm. 

Alarm Relays

The SPST 24V 500mA relay contacts are normally open and 

will close on alarm conditions when the alarms are enabled. 

Relay  contacts  are  always  functional  when  the  alarm  is 

enabled.  The  relay  will  close  only  during  minimum  and 

maximum alarm conditions. Two pen models have 2 sets 

of relay contacts which operate independently. Relays will 

not operate under battery only power conditions.

 

Dip Switch Setup

To set up the

 KT8 

recorder for your specific application, you 

might need to change some of the “Dip Switches”.

 

The Dip 

Switches are located on the back of the unit. (Figure 3)
Use  a  pen  or  small  screw  driver  to  flip  the  switches. 

Remember  to  install  the  correct  chart  to  match  the 

corresponding Dip Switch setting.

 

1

2

3

4

5

6

7

8

Slide toggle up for “on” position

Slide toggle up for “on” position

Dip Switch 1

7 & 8 are  

not active on  

this recorder.

Slide toggle down for “off” position

KT8 Dip Switch Setting  Up=On  Down=Off

Switches 1–6 For Model

1

2

3

4

5

6

7-Day

All Models

D

D

24-Hour

All Models

D

U

31-Day

All Models

U

U

Fahrenheit

All Models

D

Celsius

All Models

U

-50 to 50° F/C All Models

D

D

U

0° to 100°F/C All Models

D

D

D

0° to 250°F/C All Models

D

U

U

0° to 500°F/C KT802/03/

05/55/56

U

D

D

0 to 150° F/C KT802/03/

05/55/56

D

U

D

Remember to install the correct chart to match corresponding switch setting

DICKSON 

 

Addison, IL

 

• 

www.DicksonData.com

KT8 Dip Switch Setting  Up=On  Down=Off

Switches 1–6 Model

1

2

3

4

5

6

7-Day

KT804

D

D

24-Hour

KT804

D

U

Fahrenheit

KT804

D

Celsius

KT804

U

0 to 1000° F/C KT804

D

D

D

0 to 2000° F

KT804

D

D

U

-20° to +20°F/C

KT804

D

U

D

+20° to +100°F/C KT804

D

U

U

+50 to 150° F/C KT804

U

D

D

Remember to install the correct chart to match corresponding switch setting

DICKSON 

 

Addison, IL

 • 

www.DicksonData.com

Recording Time

All the 

KT8

 recorders feature user selectable 

recording times.

24-hour  all models

#1 down

#2 up

7-day  all models

#1 down

#2 down

31-day

 

KT802/03/56

#1 up

#2 up

°F/C 

You can record in °F or °C with the 

KT8

 by using dip 

switch #3.
°F

#3 down

°C

#3 up

Note: 

Remember to install the correct chart to match 

corresponding switch setting.