OPERATOR’S MANUAL
Instalab 700 NIR Product Analyzer
11001-1574-201107
ANALYZING PRODUCT 59
ANALYZING PRODUCT
NOTE:
Daily and weekly maintenance
routines are recommended for
optimum unit performance.
Refer to the Maintenance
section for details.
The Main Menu screen appears after all Startup screens have loaded. The
“Home” menu is the main screen that all other screens are accessed.
At the Main Menu screen, three functions are available:
1. To analyze grain
2. To access test results
3. To setup/customize the unit
IMPORTANT: The unit is designed to remain on and not be turned off.
If turned off, the unit should be powered on for 3 hours
before measurements are taken. It is recommended that
NO analysis is performed until the temperature bar is
consistently green. Stabilization can take a minimum of
30 minutes but is recommended that the unit be on for 3
hours before performing an analysis.
Figure 56
Main Menu (Home) Screen
OPTICS TEMPERATURE THERMOMETER BAR
An optics temperature thermometer displays the temperature of the optics
chamber. The bar is red until the temperature reading is /- .50
degrees C of the optimum set point of 50.00 degrees C. The bar turns green
when the temperature is /-.10 degrees C of the setpoint. A
temperature warning appears when the bar is red but analysis can still
occur. Any analysis performed when the bar is red is flagged with a symbol
(+ high temperature and - low temperature) indicating a suspect result.
Upon power up of the unit, the temperature will climb to 50.00 degrees C
and likely overshoot. Some oscillation will occur before stabilizing.
Optics Temp
Bar Status
Operating Temperature
Limits
Green
Within normal range for an/- .10 degrees C of set point
Yellow
Optics temperature is /- .50 degrees C of set point. Grain
analysis is not inhibited but not recommended.
Red
Optics temperature is greater than +/- .50 degrees C of set point.
Prediction analysis is not prohibited but not recommended. Data
collection is prohibited.
DICKEY-john Corporation
INSTALAB 700
DICKEY-john Corporation
INSTALAB 700
Analyze
Setup
Results
Power
Off
Instrument
Information
Green (normal range)
Red (out of limits;
triggers alarm)
Yellow (outside of normal
range; no alarm
occurs)
Optics Temperature (49.90 degrees C)
15
25
35
45
50
55
65
Содержание Instalab 700 NIR
Страница 4: ...OPERATOR S MANUAL Instalab 700 NIR Product Analyzer 11001 1574 201107 IV...
Страница 6: ...OPERATOR S MANUAL Instalab 700 NIR Product Analyzer 11001 1574 201107 2 SAFETY NOTICES...
Страница 12: ...OPERATOR S MANUAL Instalab 700 NIR Product Analyzer 11001 1574 201107 8 INTRODUCTION...
Страница 38: ...OPERATOR S MANUAL Instalab 700 NIR Product Analyzer 11001 1574 201107 34 NAVIGATION...
Страница 70: ...OPERATOR S MANUAL Instalab 700 NIR Product Analyzer 11001 1574 201107 66 ANALYZING PRODUCT...
Страница 74: ...OPERATOR S MANUAL Instalab 700 NIR Product Analyzer 11001 1574 201107 70 RESULTS...
Страница 78: ...OPERATOR S MANUAL Instalab 700 NIR Product Analyzer 11001 1574 201107 74 DATABASE STORAGE RECORD...
Страница 86: ...OPERATOR S MANUAL Instalab 700 NIR Product Analyzer 11001 1574 201107 82 DIAGNOSTICS...
Страница 96: ...OPERATOR S MANUAL Instalab 700 NIR Product Analyzer 11001 1574 201107 92 MAINTENANCE...
Страница 104: ...OPERATOR S MANUAL Instalab 700 NIR Product Analyzer 11001 1574 201107 100 APPENDIX B...
Страница 108: ...OPERATOR S MANUAL Instalab 700 NIR Product Analyzer 11001 1574 201107 104 APPENDIX D...
Страница 110: ...OPERATOR S MANUAL Instalab 700 NIR Product Analyzer 11001 1574 201107 106 APPENDIX E...