Tab. 4: Immunity to proximity fields from RF wireless communication equipments
The
PulsePen
is intended to be used in an electromagnetic environment in which radiated RF disturbances are
controlled. The user of the
PulsePen
can help prevent electromagnetic interference by maintaining a minimum
distance between portable / mobile RF communication equipments (transmitters) and the
PulsePen
as
recommended below, according to the maximum output power of the communications equipment.
Test Frequency
(MHz)
Band
(MHz)
Max Power
(W)
Distance
(m)
Immunity Test Level
(V/m)
385
380 - 390
1.8
0.3
27
450
430 - 470
2
0.3
28
710, 745, 780
704 - 787
0.2
0.3
9
810, 870, 930
800 - 960
2
0.3
28
1720, 1845, 1970
1700 - 1990
2
0.3
28
2450
2400 - 2570
2
0.3
28
5240, 5500, 5785
5100 - 5800
0.2
0.3
9
Minimum separation distances for higher immunity test levels shall be calculated as follows:
E= (6/d )
√
P
Where P is the maximum power (W), d is the minimum separation distance (m), E is the immunity test level (V/m).
!