Tab. 3: Electromagnetic Immunity
Power frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
30 A/m
IEC 60601-1-2
Test level
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
The
PulsePen
is suitable for use in the specified electromagnetic environment. The user of the
PulsePen
should
assure that it is used in an electromagnetic environment as described below.
Immunity
Test
IEC 60601-1-2
Test level
Compliance
level
Electromagnetic Environment
Radiated RF
IEC 61000-4-3
10V/m
80 MHz - 2.7 GHz
10V/m
Levels characteristic of a typical location in a
typical commercial or hospital environment.
Conducted RF
I.C. 61000-4-6
3V
150 KHz - 80 MHz
6V ISM bands and
amateur radio
150 KHz - 80 MHz
Not Applicable
Levels characteristic of a typical location in a
typical commercial or hospital environment.
Interference may occur in the vicinity of equipment marked with the following symbol:
&
!