Errors with integrated conductivty electrodes
In case of problem with one integrated conductivity electrode, the controller continues to run, does not
display the corresponding conductivity, but instead one error code on the « DETAILS » screen.
8.3
O
THER PROBLEMS
Error code
CAUSE
SUGGESTED ACTION
1
The temperature sensor is out of order Change the conductivity electrode (951107)
2
The conductivity sensor is out of order
Change the conductivity electrode (951107)
3
The 2 preceeding errors occur
Change the conductivity electrode (951107)
PROBLEMS
POTENTIAL CAUSES
SUGGESTED ACTION
-
The screen is totally black
-
Power supply is defective
-
Connect and switch ON the
water purifier
-
Power load is too high (pump,
valves or power supply unit in
short circuit)
-
Screen ON
-
Impossible to produce
water
-
Pump stopped
-
« NO FEED » is displayed
-
The water inlet is blocked or its
pressure is too low
-
Check the water supply
-
The water inlet tubing is bent of
obstructed
-
Check the water inlet tubing
-
Screen ON
-
Low flow of production
water
-
The pump is functional
-
Clogging of pre-treatment cartridge
-
Replace the post and pre-
treatment filters
-
Low flow of water supply
-
Increase the flow of water
supply
-
Colmatage de la membrane d’osmose
inverse
-
Change the membrane
-
Screen ON
-
Lack of pressure
-
Low flow of production
water
-
Pressurized water tank valve closed
-
Open the valve of the water
tank
-
Quantity of purified water requested
above the capacity the system
-
Wait until the pressurized
water tank is full
-
Lack of air pressure in the pressurized
water tank
-
Readjust the air pressure in the
pressurized water tank to 0,7
Bar
Problem with
the conductivity
electrode
DETAILS
INLET
MEMBRANE
3.3
2.0
PRESSURE (BAR)
EC (US/CM)
PRODUCTION
MEMBRANE
ERROR 1
20