Op
erator’s manual
version: 2021.06
24
WATER PURIFIER START
-
UP
1.
Connect all the tubings to the water purifier.
2.
Check all hydraulic connections.
3.
Close the storage tank valve.
4.
Open the water purifier exit hand valve.
5.
Open the tap water supply valve.
6.
Connect the power supply cable and switch ON the water purifier.
7.
The control unit goes through the following phases:
1.0
US/CM
120
FLUSH
The water purifier washes the membrane.
The conductivity may be different from 1.0
1.0
US/CM
59
PURGE
The water purifier rejects the stagnant water.
The conductivity may be different from 1.0
1.0
US/CM
PROD
The pump is running and the water purifier is
producing purified water.
The conductivity may be different from 1.0
8.
Wait until the system has produced 5 liters of water.
9.
The displayed production conductivity shall be between
0 and 0.1 µS/cm
. close the outlet water valve.
10.
Open the valve of the pressurized water tank and wait until it is full: the pump must stop and the
controller must display
again “READY”.
11.
The water purifier is ready to use.
An error message may appear on the screen when switching on the device, as long as the
produced water does not come out of the tap. After a few seconds the error message will
disapped, and the conductivity value will appeared on the screnn.