Step by step instructions
DIA QS / BIL QS
Quality assurance
Photometer testing with control caps
Diaglobal GmbH · Köpenicker Straße 325 · 12555 Berlin · +49 (0)30 6576 2597 · [email protected] · www.diaglobal.de 2021-01
1. 20 control caps with lyophilized
control serum
DIA QS: Glucose, Lactate,
Triglycerides, Cholesterol, HDL-
Cholesterol, Uric acid, Bilirubin (adults)
BIL QS: Bilirubin (adults), Bilirubin
N (newborns)
2. Screw the control cap onto a
cuvette of the test you want to
check
Mix well
Now the cuvette contains a sample
with a known concentration
3. Leave the cuvette for 1
minute
4. Switch photometer on with
ON/ENTER key
Wait for device check, confirm
with ON/ENTER
Select the required test,
confirm with ON/ENTER
6. Replace control cap with
starter cap
ON
ENTER
GLU
8. First press ON/ENTER, then
insert cuvette into photometer
Wait for measured value
Compare the measured value with
the target value on the package
insert
7. Turn cuvette upside down
several times
5. Zero point adjustment:
Insert cuvette with sample
(Fig. 3) into photometer, zero
point is stored in memory
Remove cuvette after signal
tone