Step by step instructions
BIL / BIL N (BIL 142)
Additionally required: Mini centrifuge
Diaglobal GmbH · Köpenicker Straße 325 · 12555 Berlin · +49 (0)30 6576 2597 · [email protected] · www.diaglobal.de 2021-01
1. Insert capillary with sample
(serum/plasma) into cuvette
BIL: 100 µL (adults)
BIL N: 20 µL (newborns)
Sampling of BIL N see next
page
2. Eject sample several times
with micropipetter into
cuvette
3. Screw cap on
Turn cuvette upside down
several times
5. Zero point adjustment:
Insert cuvette with sample
(Fig. 3) into photometer, zero
point is stored in memory
Remove cuvette after signal
tone
4. Switch photometer on with
ON/ENTER key
Wait for device check, confirm
with ON/ENTER
Select the required test,
confirm with ON/ENTER
6. Replace screw cap with
starter cap
ON
ENTER
ON
ENTER
BIL
Sample 01
insert
9. Time is displayed, wait for
measured value
7. Turn cuvette upside down
several times
8. First press ON/ENTER
Then insert cuvette into
photometer
ON
ENTER
BIL
Sample 01 00:30
Measuring
BIL
Sample 01
ENTER