Step by step instructions
GOT 442 / GPT 442
Number of samples per series: Up to 6 samples at the same time
Additionally required: Dry block thermostat (30 minutes preheated)
Diaglobal GmbH · Köpenicker Straße 325 · 12555 Berlin · +49 (0)30 6576 2597 · [email protected] · www.diaglobal.de 2021-01
1.
Incubate cuvettes in preheated
dry block thermostat for 7 minutes
2.
Transfer 50 µL of the sample
(serum/plasma) with an end-to-end
capillary into each cuvette
Do not mix yet!
3.
Screw starter cap on and mix well
Then insert the cuvette immediately
into the dry block thermostat
5.
Start measurement with ON/ENTER
Time (1 minute) counts backwards. All
cuvettes remain in the dry block
thermostat during this time
Double signal tone: The M1
measurement starts in 30 seconds!
4.
Switch photometer on with
ON/ENTER key
Wait for device check, confirm with
ON/ENTER
Select GOT or GPT, confirm with
ON/ENTER
6.
Note: When inserting the cuvette
into the photometer, make sure that the
capillary is positioned at the side of the
cuvette wall and aligns with the black
arrow on the front foil
This ensures that the measurement is
not disturbed by the capillary.
ON
ENTER
ON
ENTER
GOT
GOT Measurement
START
9.
All measured values can now be
displayed one after the other by
pressing the right arrow key
7.
Follow
the display, insert sample 1 into
the photometer, „Measuring“ is displayed,
wait for 10 seconds
Then remove the cuvette and incubate it
again
Proceed in the same way with all other
cuvettes in the correct order
Then press ON/ENTER
8.
Time (10 minutes) counts
backwards. All cuvettes remain in the
dry block thermostat during this time
Double signal tone: The M2
measurement starts in 30 seconds!
For the M2 measurement follow the
display in the same way as discribed in
Fig. 7
ON
ENTER
M2 GOT
Sample 09:30
incubate
M1 GOT
Samples 00:30
incubate
ON
ENTER
Result GOT
Sample 01
40 U/L
ON
ENTER
M1 GOT
Sample 02 00:03
insert or
ENTER